Skip to content

EndoWrist 471093Recalls & reported problems

Intuitive Surgical, Inc. · PROGRASP FORCEPS · Class 2

All models of EndoWrist (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Intuitive EndoWrist ProGrasp forceps for da Vinci systems appears in 8,631 adverse event reports, including 171 injuries and no deaths. The reports mostly describe frayed, torn or broken cables and parts that split or fragment. One open recall, from December 2024, concerns frayed or broken grip cables on reusable instruments with jaws.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
8,581
Deaths reported
0
Injuries reported
173
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8,581 adverse event reports that list this exact device (DI 00886874119785), received between Dec 10, 2020 and Sep 30, 2026 — classified as 173 injuries, 8,408 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (4,852 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 18 reportsDec 2023: 12 reportsJan 2024: 13 reportsFeb 2024: 19 reportsMar 2024: 25 reportsApr 2024: 22 reportsMay 2024: 20 reportsJun 2024: 25 reportsJul 2024: 29 reportsAug 2024: 17 reportsSep 2024: 26 reportsOct 2024: 28 reportsNov 2024: 13 reportsDec 2024: 16 reportsJan 2025: 16 reportsFeb 2025: 52 reportsMar 2025: 1,257 reportsApr 2025: 568 reportsMay 2025: 670 reportsJun 2025: 586 reportsJul 2025: 517 reportsAug 2025: 397 reportsSep 2025: 289 reportsOct 2025: 240 reportsNov 2025: 341 reportsDec 2025: 325 reportsJan 2026: 304 reportsFeb 2026: 1,722 reportsMar 2026: 353 reportsApr 2026: 184 reportsMay 2026: 98 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 44 reportsSep 2026: 58 reportsOct 2026: 0 reports2024202520261,722
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Open, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0874-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn4,852
  • Material Frayed1,829
  • Material Fragmentation572
  • Detachment of Device or Device Component394
  • No Apparent Adverse Event289
  • Break187
  • Unintended Movement172
  • Unintended System Motion97

Patient problems

  • No Clinical Signs, Symptoms or Conditions6,312
  • Insufficient Information2,191
  • Foreign Body In Patient67
  • Unspecified Tissue Injury25
  • Hemorrhage/Blood Loss/Bleeding5
  • Bowel Perforation2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bruise/Contusion1

Recent adverse event reports

The 25 most recent of 8,581 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954554 ↗
  • MalfunctionSep 30, 2026Entrapment of Device; Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954548 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954386 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50954331 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954322 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954317 ↗
  • MalfunctionSep 30, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50954308 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50954304 ↗
  • MalfunctionSep 30, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953290 ↗
  • MalfunctionSep 30, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949834 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    It was reported that there was a broken cable with a da Vinci product. The customer reported event has no report of patient injury.

    Patient problems: Insufficient Information

    MAUDE report 50942365 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939696 ↗
  • MalfunctionSep 29, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50926986 ↗
  • MalfunctionSep 28, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926099 ↗
  • MalfunctionSep 28, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926048 ↗
  • MalfunctionSep 28, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925982 ↗
  • MalfunctionSep 28, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50925336 ↗
  • MalfunctionSep 28, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925036 ↗
  • MalfunctionSep 28, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925023 ↗
  • MalfunctionSep 28, 2026Material Twisted/Bent

    Patient problems: Insufficient Information

    MAUDE report 50924713 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50905849 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50905839 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905832 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905800 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    Patient problems: Insufficient Information

    MAUDE report 50905762 ↗

Other System, surgical, computer controlled instrument devices

All System, surgical, computer controlled instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page