NONE 470500Recalls & reported problems
Intuitive Surgical, Inc. · SEAL · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This da Vinci Universal Seal, a part used in robotic surgery, has 140 adverse event reports, with no deaths and 56 injuries. Many describe pieces breaking off or detaching during procedures, sometimes falling inside the patient and being retrieved during the same operation, as well as gas leaks. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 127
- Deaths reported
- 0
- Injuries reported
- 51
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 127 adverse event reports that list this exact device (DI 00886874116401), received between Apr 28, 2023 and Jun 11, 2026 — classified as 51 injuries, 76 malfunctions.
The most frequently reported specific device problem is “Break” (47 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break47
- Gas/Air Leak31
- Detachment of Device or Device Component25
- Material Fragmentation24
- Material Split, Cut or Torn9
- Material Separation3
- Appropriate Device Problem Term/Code Not Available1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions102
- Foreign Body In Patient21
- Insufficient Information5
Recent adverse event reports
The 25 most recent of 127 reports, in the reporter's own words as published by the FDA.
InjuryJun 11, 2026Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ABDOMINOPERINEAL RESECTION PROCEDURE, A SMALL BLACK PIECE FROM THE UNIVERSAL SEAL WAS IDENTIFIED IN THE PATIENT'S PELVIS. THE FRAGMENT HAD REPORTEDLY FALLEN INSIDE THE PATIENT AND WAS RETRIEVED DURING THE SAME PROCEDURE. THE CUSTOMER STATED THAT NO DAMAGE TO THE UNIVERSAL SEAL WAS IDENTIFIED PRIOR TO USE. TOWARDS THE END OF THE PROCEDURE, A PIECE OF PLASTIC WAS OBSERVED IN THE PATIENT'S PELVIS. A LAPAROSCOPIC GRASPER WAS ALREADY INSERTED THROUGH ONE OF THE PORTS AT THE TIME. THE SURGEON SUSPECTS THE FRAGMENT MIGHT HAVE BROKEN OFF DURING INSERTION OR WITHDRAWAL OF EITHER A ROBOTIC INSTRUMENT FROM PORT #3 OR #4 OR A LAPAROSCOPIC GRASPER INSTRUMENT. THE FRAGMENTS WERE REMOVED USING A LAPAROSCOPIC GRASPER. AN INSPECTION OF THE UNIVERSAL SEAL CONFIRMED THAT ONLY A SINGLE PIECE HAD BROKEN OFF. NO POSTOPERATIVE IMAGING, SUCH AS X-RAY OR ULTRASOUND, WAS PERFORMED TO CHECK FOR RETAINED FRAGMENTS. THE PATIENT HAS NOT RETURNED WITH ANY POSTOPERATIVE COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25477587 ↗MalfunctionJun 8, 2026
THE CASE WAS A ROBOTIC LAPAROSCOPIC HYSTERECTOMY BILATERAL SALPINGECTOMY. WHEN THE OBTURATOR WAS PUT THROUGH TROCAR, DR. [REDACTED] SAW THAT A SMALL PORTION OF THE SEAL HAD BROKEN OFF INTO THE PATIENT.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 25425813 ↗InjuryMay 18, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER SUSPECTED THAT A PIECE OF RUBBER BROKE OFF FROM ONE OF TWO UNIVERSAL SEALS AND FELL INTO THE PATIENT'S BODY CAVITY. THE CUSTOMER RETRIEVED THE FRAGMENT USING A ROBOTIC INSTRUMENT DURING THE SAME PROCEDURE. NO POST-OPERATIVE DIAGNOSTIC STUDIES WERE PERFORMED AND NO ADDITIONAL SURGERIES WERE REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25204089 ↗MalfunctionMay 14, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ISUFFLATION SITE CONNECTOR BROKE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CUSTOMER NOTED IT BROKE DURING PROCEDURE. THE PART OF THE SEAL THAT BROKE DID NOT EFFECT THE INSUFFLATION WITH NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176807 ↗MalfunctionMay 14, 2026
WHILE ON THE ROBOT FOR ROBOTIC LAPAROSCOPIC CHOLECYSTECTOMY, AN INSTRUMENT WAS PLACED IN THE TROCAR FORCEFULLY AND CUT THE RUBBER SEAL OFF THE TROCAR CAP. IT WAS FOUND IN THE ABDOMEN DURING THE PROCEDURE. CONFIRMED AT THE END OF THE CASE WHERE IT CAME FROM AND THAT WE HAD ALL THE PIECES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25168185 ↗MalfunctionMay 7, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED TRANSTHORACIC ESOPHAGECTOMY NECK ANASTOMOSIS SURGICAL PROCEDURE, THE AIR INSUFFLATION PORT BROKE OFF WHILE THE INSUFFLATION TUBING WAS IN USE. THE USER CONTINUED WITH THE PROCEDURE AS PLANNED. NO FRAGMENT FELL INTO THE PATIENT. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25117012 ↗InjuryApr 22, 2026Material Separation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY PROCEDURE, A BLACK FRAGMENT FROM A UNIVERSAL SEAL CAME OFF AND FELL INTO THE PATIENT. THE SURGEON WAS ABOUT ¾ COMPLETED WITH THE PROCEDURE WHEN A PIECE OF THE UNIVERSAL SEAL WAS OBSERVED TO HAVE FALLEN OFF AND WAS VIEWABLE IN THE PATIENT ANATOMY. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PIECE WITH NO INJURY OR HARM TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: DURING THE PROCEDURE, THE UNIVERSAL SEAL WAS INSPECTED AND PLACED ON THE TROCAR, WITH NO ABNORMALITIES OBSERVED AT THAT TIME. THE TROCAR, POSITIONED IN ARM #4 (THE SURGEON'S RIGHT ARM), WAS EXCHANGED MULTIPLE TIMES FOR SCISSORS, CLIPS, AND A NEEDLE HOLDER INSTRUMENTS. IT IS UNCLEAR EXACTLY WHEN THE FRAGMENT FELL INTO THE PATIENT. THE UNIVERSAL SEAL WAS IN USE FOR APPROXIMATELY 2 HOURS. THE SURGEON IS UNCERTAIN WHAT CAUSED THE PIECE TO FALL OFF. THE PORT REMAINED IN PLACE UNTIL THE END OF THE PROCEDURE AND FUNCTIONED PROPERLY THROUGHOUT THE PROCEDURE, WITH NO INDICATION OF A SEAL MALFUNCTION. THE FRAGMENT WAS SUBSEQUENTLY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24955873 ↗MalfunctionApr 21, 2026Break; Gas/Air Leak
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, AFTER 4 HOURS OF DOCKING, THE SEAL SPONTANEOUSLY BROKE AT THE THREADED PART. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE DAMAGE RESULTED IN A RAPID LOSS OF INSUFFLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24944947 ↗InjuryApr 14, 2026Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LIVER RESECTION PROCEDURE, THE RUBBER COMPONENT OF THE UNIVERSAL SEAL BROKE AND A FRAGMENT FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED DURING A DIFFERENT PROCEDURE. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP TO OBTAIN ADDITIONAL INFORMATION; HOWEVER, NO FURTHER DETAILS WERE AVAILABLE REGARDING THE REPORTED EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 24873029 ↗MalfunctionApr 13, 2026Material Fragmentation; Gas/Air Leak
IT WAS REPORTED THAT DURING A DA VINCI ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL HAD AIR LEAK. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: IT WAS UNKNOWN IF THE EVENT RESULTED IN SUDDEN/RAPID OF INSUFFLATION. A BACKUP UNIVERSAL SEAL WAS USED FOR THE PROCEDURE WITH NO PATIENT INJURY OR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24858918 ↗MalfunctionApr 11, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSUFFLATION TUBE BROKE WHEN IT WAS ATTACHED. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24851294 ↗MalfunctionApr 11, 2026Break
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24848775 ↗MalfunctionApr 8, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL ACCESSORY BROKE OFF INTO THE INSUFFLATION. THE PROCEDURE WAS COMPLETED WITH NO REPORTED PATIENT INJURY. INTUITIVE FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: DURING THE PROCEDURE, A PIECE OF THE SEALING CAP BROKE OFF TWICE, CAUSING A RAPID LOSS OF INSUFFLATION. IN BOTH INSTANCES, THE BROKEN PIECE REMAINED INSIDE THE INSUFFLATION TUBING AND DID NOT ENTER THE PATIENT, WITH NO MISSING FRAGMENTS REPORTED. THE FIRST OCCURRENCE WAS IMMEDIATELY NOTICED UPON ATTACHING THE TUBING, AND THE SECOND WAS IDENTIFIED DURING THE PROCEDURE BY THE WHOOSHING SOUND. BOTH ISSUES WERE RESOLVED BY REPLACING THE SEALING CAP OR SWITCHING THE TUBING TO ANOTHER PORT, RESULTING IN MINIMAL PROCEDURAL DELAY (LESS THAN 3 MINUTES). A PHOTO OF THE PIECE FROM THE FIRST INCIDENT WAS AVAILABLE, BUT NO MATERIALS REMAIN FROM THE SECOND INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24816036 ↗MalfunctionApr 3, 2026Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL LUER PORT WHERE THE INSUFFLATION TUBE CONNECTS WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781661 ↗MalfunctionApr 3, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL PORT HAD A MALFUNCTION AND HAD A DAMAGED PART. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24780209 ↗MalfunctionMar 26, 2026Break; Gas/Air Leak
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL BROKE OFF AND CAUSED LOSS OF INSUFFLATION. THE CUSTOMER USED A NEW UNIVERSAL SEAL TO CONTINUE WITH THE SURGICAL PROCEDURE. THERE WAS NO PATIENT HARM. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24708369 ↗MalfunctionMar 20, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SEAL ACCESSORY CONNECTOR BROKE OFF. THE CANNULA WAS REMOVED AND REPLACED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED ADDITIONAL INFORMATION: THERE WERE NO MISSING FRAGMENTS, AND THE ISSUE WAS RESOLVED BY REPLACING THE UNIVERSAL SEAL ACCESSORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654672 ↗InjuryMar 17, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED THORACIC PROCEDURE, A FRAGMENT FROM A UNIVERSAL SEAL BROKE OFF AND FELL INSIDE THE PATIENT. ALL FRAGMENTS WERE SUCCESSFULLY RETRIEVED DURING THE SAME PROCEDURE WHICH WAS COMPLETED ROBOTICALLY. INTUITIVE SURGICAL, INC. (ISI) HAS ATTEMPTED TO GATHER ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24624519 ↗InjuryMar 9, 2026Material Separation
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24546830 ↗MalfunctionJan 12, 2026Break; Gas/Air Leak
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE PORT BROKE OFF WHEN INSUFFLATION WAS ATTACHED FROM THE UNIVERSAL SEAL. THE PRODUCT WAS DISCARDED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24047154 ↗MalfunctionJan 9, 2026Break
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24030135 ↗MalfunctionJan 7, 2026Gas/Air Leak
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE AIR LEAKED FROM THE CANNULA SEA WHILE ATTEMPTING PNEUMOPERITONEUM. THE PROCEDURE WAS COMPLETED WITH A BACKUP CANNULA SEAL, WITH NO REPORTS OF PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION ON 26-DEC-2025: THE PORT OR LEVER WAS REPORTEDLY BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23995424 ↗MalfunctionJan 7, 2026Gas/Air Leak
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23995382 ↗MalfunctionJan 7, 2026Gas/Air Leak
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23993720 ↗MalfunctionDec 5, 2025Break; Gas/Air Leak; Break; Gas/Air Leak
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23730243 ↗
FDA clearance and classification
The FDA regulates product code GCJ (laparoscope, general & plastic surgery) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the NONE 470500
Has the NONE 470500 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00886874116401), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NONE 470500?
The FDA's MAUDE database lists 127 adverse event reports for this device, received between Apr 28, 2023 and Jun 11, 2026. In those reports, the most common specific device problems are “Break” (47), “Gas/Air Leak” (31) and “Detachment of Device or Device Component” (25); the most common patient problem is “Foreign Body In Patient” (21). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NONE 470500?
None of the reports are classified as deaths, 51 as injuries and 76 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the NONE 470500, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand NONE, model 470500). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,913 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,857 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,068 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall945 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery850 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery770 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall694 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery569 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page