None 470066Recalls & reported problems
Intuitive Surgical, Inc. · ENDOSCOPE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00886874120293), received between Dec 2, 2025 and Sep 21, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Image Orientation Incorrect” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Image Orientation Incorrect2
- Material Integrity Problem1
- No Device Output1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Image Orientation Incorrect
It was reported that during a da Vinci-assisted surgical procedure, while targeting the customer found that the image angle shown by the 0 degree Endoscope did not match the horizon indicator shown on the system. The customer reseated the Endoscope and the issue was resolved. The procedure was completed with no reported injury.ISI followed up with the initial reporter and obtained the following additional information: The customer did not observe any damage or missing screws on the Endoscope base. The Endoscope was not manually rotated prior to the issue. The endoscope will be returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50843379 ↗MalfunctionMar 30, 2026Output Problem
IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM ENDOSCOPE, 0° HAD A FAULT 45312 "LOSS OF THE RIGHT EYE". THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: NO FRAGMENTS FALL INTO THE PATIENT'S BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24736203 ↗MalfunctionFeb 19, 2026Material Integrity Problem
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 0° ENDOSCOPE WAS FOUND TO BE BENT AT THE DISTAL TIP. THE PROCEDURE WAS COMPLETED WITH NO REPORTED PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 24385682 ↗MalfunctionJan 14, 2026Image Orientation Incorrect
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24071756 ↗MalfunctionDec 2, 2025No Device Output; No Device Output
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ENDOSCOPE FAILED NUMEROUS ATTEMPTS TO ENGAGE WITH THE ARM. THE CUSTOMER RESEATED THE DRAPE, TRIED A NEW DRAPE, REPROCESSED, AND ATTEMPTED TO USE A DIFFERENT ROBOT, BUT EACH TIME THE ENDOSCOPE FAILED TO ENGAGE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694690 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page