Skip to content

PRECICE® Plating System LP4.5-DRF40Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 0-Hole · Class 2 · Implant

All models of PRECICE Plating System (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1325-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates