XLIF 6940430Recalls & reported problems

Nuvasive, Inc. · XLIF Pituitary, Medium · Class 1 · Implant

All models of XLIF (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00887517452771), received between Jun 3, 2022 and Feb 24, 2026 — classified as 20 malfunctions.

The most frequently reported specific device problem is “Fracture” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture20

Patient problems

  • No Clinical Signs, Symptoms or Conditions20

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Fracture

    THE CLOSING/ARTICULATING TIP OF THE PITUITARY BROKE DURING DISCECTOMY. THE TIP WAS RETRIEVED, AND NO HARM WAS DONE TO THE PATIENT. THE PROCEDURE WAS COMPLETED AS SCHEDULED REPLACING THE TLIF INTERBODY CAGE WITH A COHERE ALIF INTERBODY CAGE. THE PATIENT WAS DISCHARGED NORMALLY AND IS REPORTED TO BE DOING VERY WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24433244 ↗
  • MalfunctionFeb 20, 2026Fracture

    IT WAS REPORTED THAT THE XLIF PITUITARY BROKE DURING SURGERY. IT WAS CONFIRMED NO PIECES OF THE INSTRUMENT FELL INTO THE PATIENT. THERE WAS NO ADVERSE EFFECT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405660 ↗
  • MalfunctionFeb 4, 2026Fracture

    THE TIP OF MEDIUM SIZED XLIF PITUITARY BROKE OFF DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260970 ↗
  • MalfunctionFeb 3, 2026Fracture

    IT WAS REPORTED THAT THE HEAD OF PITUITARY BROKE OFF DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251439 ↗
  • MalfunctionSep 30, 2025Fracture; Fracture

    THE DEVICE TIP WAS FRACTURED DURING SURGERY. THE TIP WAS RETRIEVED DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23181318 ↗
  • MalfunctionSep 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF A XLIF PITUITARY BROKE OFF DURING SURGERY AND WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23018048 ↗
  • MalfunctionAug 25, 2025Fracture; Fracture

    THE PITUITARY MOUTH BROKE OFF WHILE TRYING TO DO DISCECTOMY. ALL PARTS WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882803 ↗
  • MalfunctionApr 17, 2025Fracture; Fracture

    FRACTURED TIP WOULD NOT FUNCTION DURING EXTREME LATERAL INTERBODY FUSION, NO FRAGMENTS IN PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21861943 ↗
  • MalfunctionDec 5, 2024Fracture; Fracture

    DURING A PROCEDURE, THE DISTAL ARTICULATING TIP OF A PITUITARY RONGEUR FRACTURED OFF. THE PIECE WAS RETRIEVED AND WAS NOT RETAINED IN THE PATIENT. NO FURTHER PATIENT OR PROCEDURE IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20856139 ↗
  • MalfunctionJul 17, 2024Fracture; Fracture

    DURING AN EXTREME LATERAL INTERBODY FUSION PROCEDURE, THE DISTAL ARTICULATING TIP OF THE PITUITARY RONGEUR FRACTURED OFF. THE PIECE WAS RETRIEVED AND WAS NOT RETAINED IN THE PATIENT. NO FURTHER PATIENT OR PROCEDURE IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19767479 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    DURING A PROCEDURE, THE DISTAL ARTICULATING TIP OF THE PITUITARY RONGEUR FRACTURED OFF. THE PIECE WAS RETRIEVED AND WAS NOT RETAINED IN THE PATIENT. NO FURTHER PATIENT OR PROCEDURE IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19519725 ↗
  • MalfunctionMar 14, 2024Fracture; Fracture

    IT WAS REPORTED THE TIP OF PITUITARY RONGEUR FRACTURED OFF IN THE DISC SPACE DURING A PROCEDURE. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENT, USER, OR PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18909503 ↗
  • MalfunctionDec 6, 2023Fracture; Fracture

    IT WAS REPORTED THE PITUITARY TIP FRACTURED OFF DURING A PROCEDURE WHILE REMOVING AN OSTEOPHYTE. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENT, USER, OR PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18275406 ↗
  • MalfunctionNov 19, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE JAW FRACTURED OFF DURING PROCEDURE. THE BROKEN PART WAS RECOVERED AND DISCARDED. THERE WAS NO REPORT OF ADVERSE PATIENT OR PROCEDURE IMPACT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18167306 ↗
  • MalfunctionOct 23, 2023Fracture; Fracture

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993007 ↗
  • MalfunctionMay 17, 2023Fracture; Fracture

    DURING A SPINAL PROCEDURE THE TEETH WERE BROKEN AND EVERYTHING WAS REMOVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16946420 ↗
  • MalfunctionApr 4, 2023Fracture; Fracture

    DURING A SPINAL PROCEDURE THE DISTAL GRABBING FEATURE BROKE AND THE SEPARATED PIECE WAS RETRIEVED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16675183 ↗
  • MalfunctionFeb 27, 2023Fracture; Fracture

    ON (B)(6) DURING A SPINAL PROCEDURE IT WAS REPORTED THAT THE TOP OF THE PITUITARY BROKE OFF. ALL PIECES WERE RETRIEVED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16448953 ↗
  • MalfunctionNov 22, 2022Fracture; Fracture

    ON (B)(6) 2022 A PATIENT UNDERWENT A BILATERAL POSTERIOR LATERAL INTERBODY FUSION REVISION PROCEDURE FROM L5-S1 TO ADDRESS RETROPULSION. THE DEVICES BEING REVISED WERE NOT NUVASIVE PRODUCTS. WHILE REMOVING THE CAGE THE UPPER DISTAL JAW OF TWO DEVICES FRACTURED. THE FRACTURED PIECES WERE REMOVED FROM THE PATIENT. THERE WERE NO INJURIES OR ADVERSE PATIENT CONSEQUENCES AND THE SURGERY WAS COMPLETED WITHOUT ISSUE. REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15849175 ↗
  • MalfunctionJun 3, 2022Fracture; Fracture

    ON (B)(6) 2022 A PATIENT UNDERWENT AN EXTREME LATERAL INTERBODY FIXATION PROCEDURE. DURING THE PROCEDURE WHILE REMOVING TISSUE FROM THE DISC SPACE THE UPPER DISTAL BITING JAW FRACTURED OFF. ALL PIECES WERE RETRIEVED WITHOUT INJURY OR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14594371 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →