ReLine 11355035Recalls & reported problems
Nuvasive, Inc. · RELINE MAS Ti Rod, 5.5x35mm Lordotic · Class 2 · Implant
All models of ReLine (129)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887517504999), received between Oct 15, 2018 and Apr 15, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Migration or Expulsion of Device1
Patient problems
- No Consequences Or Impact To Patient1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 15, 2019Migration or Expulsion of Device; Migration or Expulsion of Device
ON (B)(6) 2019 PATIENT UNDERWENT AN INTERBODY FUSION PROCEDURE WITH POSTERIOR FIXATION AT S1-L2 LEVELS WITHOUT A REPORTED ISSUES. DURING A ROUTINE FOLLOW UP RADIOGRAPHS IDENTIFIED THAT A ROD PULLED OUT OF THE SET SCREW AT THE BOTTOM OF THE CONSTRUCT. ON (B)(6) 2019 A REVISION PROCEDURE TOOK PLACE, REPLACING THE LOCK SCREW AND SECURING THE ROD. NO PATIENT HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8516680 ↗InjuryOct 15, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON (B)(6) 2018, PATIENT UNDERWENT POSTERIOR FUSION AT THE L3-L4 VERTEBRAL LEVELS. SEVERAL MONTHS POST-OPERATIVE PATIENT WAS EXPERIENCING BACK PAIN AND UPON EVALUATION AN INFECTION WAS CONFIRMED. ON (B)(6) 2018, ALL THE IMPLANTED DEVICES WERE REMOVED. AS PER REPORTER ONCE THE PATIENT RECOVERS, POSTERIOR FUSION USING AUTOLOGOUS BONE TRANSPLANTATION, WITHOUT ANY IMPLANT DEVICE, WILL BE PERFORMED.
Patient problems: Pain; Unspecified Infection
MAUDE report 7967213 ↗
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page