Skip to content

ReLine 10000481Recalls & reported problems

Nuvasive, Inc. · RELINE MAS Shank Adjuster, Cannulated · Class 1

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517511782), received between Mar 13, 2026 and Mar 13, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Fracture

    A REVISION CASE, SURGEON REMOVED A 7.5X55MM RELINE SCREW AND WANTED TO REPLACE WITH AN 8.5X55 SCREW. TAPPED WITH 7.5 TAPAS THE DOC ALWAYS UNDER SIZES BY 1MM. ADVANCED THE SCREW ABOUT 40/55MM INTO THE PEDICLE AND THE SCREW BECAME SO INCREDIBLY TIGHT IT WOULDN¿T MOVE ANY FURTHER. THE SURGEON HAD NO LUCK REVERSING THE SCREW WITH THE RELINE MAS RED HANDLE, SO ATTEMPTED A SHANK DRIVER FOR MORE LEVERAGE WITH A T-HANDLE. DRIVER TIP COMPLETELY BROKE OFF INSIDE THE SHANK. WE WERE UNABLE TO BACK THE SCREW COMPLETELY OUT. ATTEMPTED TO PUT A SHORT ROD AND SET SCREW INSIDE THE TULIP TO HELICOPTER THE SCREW OUT, HAD TO SNAP THE TOWERS OFF. ROD WAS LOOSE INSIDE THE TULIP, ASSUMING THE DRIVER THAT BROKE OFF WAS PREVENTING A RIGID CONNECTION AND THEREFORE UNABLE TO LOCK THE TULIP INTO A RIGID STATE. WE SUGGESTED TO TAKE THE ENDOSCOPE AND LOOK DIRECTING INTO THE TULIP TO SEE IF WE COULD REMOVE THE BROKEN DRIVER PIECE AND THEN RETRY THE ROD AND HELICOPTER METHOD. THE SURGEON DECIDED NOT TO. A SPINE REMOVAL SET WAS AVAILABLE FOR THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592929 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page