ReLine 10000750Recalls & reported problems

Nuvasive, Inc. · RELINE MAS Final Lock Screw Driver, Long · Class 1

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00887517513717), received between May 31, 2022 and May 8, 2026 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Fracture” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture13

Patient problems

  • Foreign Body In Patient12
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Fracture

    IT WAS REPORTED THAT THE TIP OF THE RELINE MAS FINAL LOCK SCREW DRIVER WAS FRACTURED WHEN THE SURGEON USED IT FOR FINAL TIGHTENING. THE SURGEON COULD NOT REMOVE THE FRAGMENT FROM THE LOCK SCREW AND REMAINS IN THE BODY. THERE WAS NO OTHER ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25126694 ↗
  • MalfunctionAug 25, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE END OF THE RELINE DRIVER BROKE OFF DURING SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 22882723 ↗
  • MalfunctionApr 24, 2025Fracture; Fracture

    TIP OF THIS INSTRUMENT BROKE OFF INTO THE LOCK SCREW AND COLD WELDED ITSELF INTO THE LOCK SCREW AND LEFT IN-SITU.

    Patient problems: Foreign Body In Patient

    MAUDE report 21905875 ↗
  • MalfunctionDec 31, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A POSTERIOR SPINAL FUSION PROCEDURE FROM L3 TO L5, WHILE THE SURGEON WAS FINAL TIGHTENING THE LOCK SCREW AT THE RIGHT L4 LEVEL, THE TIP OF THE FINAL LOCK SCREW DRIVER FRACTURED OFF WITHIN THE LOCK SCREW. THE TIP WAS LEFT IN-SITU, EMBEDDED IN THE LOCK SCREW. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21048519 ↗
  • MalfunctionSep 26, 2024Fracture; Fracture

    IT WAS REPORTED THAT A DURING A TRANSFORAMINAL LUMBAR INTERBODY FUSION, WHILE THE SURGEON WAS FINAL TIGHTENING, THE TIP OF THE FINAL LOCK SCREW DRIVER FRACTURED OFF WITHIN THE LOCK SCREW. THE TIP WAS LEFT IN-SITU, EMBEDDED IN THE LOCK SCREW. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20311736 ↗
  • MalfunctionAug 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROCEDURE FROM L3 TO S1. DURING THE PROCEDURE, WHILE THE SURGEON WAS FINAL TIGHTENING THE LOCK SCREW ON THE LEFT S1, THE TIP OF THE FINAL LOCK SCREWDRIVER FRACTURED OFF WITHIN THE LOCK SCREW. THE TIP WAS LEFT IN-SITU, EMBEDDED IN THE LOCK SCREW. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20034201 ↗
  • MalfunctionFeb 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DRIVER TIP FRACTURED OFF. THE FRACTURED PORTION COULD NOT BE RETRIEVED AND THEREFORE REMAINS IN THE PATIENT. THERE HAS BEEN NO REPORT OF ADVERSE PATIENT IMPACT AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18628589 ↗
  • MalfunctionJun 16, 2023Fracture; Fracture

    IT WAS REPORTED THE DISTAL TIP OF THE DRIVER FRACTURED OFF WHILE FINAL TIGHTENING THE LOCK SCREW DURING A POSTERIOR PROCEDURE; THE FRACTURED TIP WAS NOT RECOVERED AND REMAINS LODGED IN THE LOCK SCREW. THERE WAS NO REPORTED ADVERSE PATIENT OR PROCEDURE IMPACT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 17148324 ↗
  • MalfunctionMay 25, 2023Fracture; Fracture

    DURING A PROCEDURE, THE DRIVER TIP FRACTURED OFF WHILE FINAL TIGHTENING THE LOCK SCREW. THE FRACTURED PORTION COULD NOT BE RETRIEVED AND THEREFORE REMAINS IN THE PATIENT; HOWEVER, THE FRACTURED PORTION WAS REPORTED TO BE LODGED WITHIN THE LOCK SCREW. THERE WAS NO DELAY IN SURGERY AND THERE HAS BEEN NO REPORT OF ADVERSE PATIENT IMPACT AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 17007269 ↗
  • MalfunctionOct 21, 2022Fracture; Fracture

    DURING A POSTERIOR FIXATION FROM LEVELS L2 TO L5 WHILE FINAL TIGHTENING THE LEFT L2 LOCK SCREW, THE TIP OF THE FINAL LOCK SCREW DRIVER FRACTURED. THE FRACTURED PIECE COULD NOT BE RETRIEVED AND AS A RESULT WAS RETAINED IN THE PATIENT. NO INJURY AND NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 15655872 ↗
  • MalfunctionSep 1, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A MINIMALLY INVASIVE POSTERIOR FIXATION PROCEDURE. DURING THE PROCEDURE, THE DISTAL TORX FEATURE OF THE FINAL TIGHTENING DRIVER FRACTURED OFF INTO THE LOCK SCREW. AFTER APPROXIMATELY 20-30 MINUTES WERE USED IN AN EFFORT TO RETRIEVE THE FRACTURED PIECE, IT WAS DETERMINED TO BE CAPTURED WITHIN THE LOCK SCREW AND UNABLE TO BE RETRIEVED. AS A RESULT, THE TIP WAS RETAINED IN THE PATIENT. NO INJURY AND NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 15341267 ↗
  • MalfunctionJun 2, 2022Fracture; Fracture

    ON (B)(6) 2022 A PATIENT UNDERWENT A SPINAL PROCEDURE. DURING THE SURGERY DURING TIGHTENING THE DEVICE TIP FRACTURED OFF IN THE RIGHT L4 SCREW AND COULD NOT BE RETRIEVED FROM THE SCREW. THE DEVICE TIP WAS SEALED THE ENGAGED SCREW BY THE ROD AND LOCK SCREW AND THE SURGERY WAS COMPLETED WITHOUT FURTHER INCIDENT. THERE WERE NO ADVERSE EFFECT TO THE PATIENT AS A CONSEQUENCE OF THE DEVICE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14577723 ↗
  • MalfunctionMay 31, 2022Fracture; Fracture

    ON (B)(6) 2022 DURING A SPINAL PROCEDURE THE TIP OF THE FINAL LOCK SCREW DRIVER WAS SHEARED OFF AND LEFT IN-SITU COLD-WELDED INTO THE LOCK SCREW.

    Patient problems: Foreign Body In Patient

    MAUDE report 14549938 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →