ReLine 13116540Recalls & reported problems
Nuvasive, Inc. · RELINE-O Screw, 6.5x40mm Traction Poly · Class 2 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887517582492), received between Aug 24, 2017 and Jan 13, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Loosening of Implant Not Related to Bone-Ingrowth1
- Migration or Expulsion of Device1
- Unintended Movement1
Patient problems
- Failure of Implant4
- No Consequences Or Impact To Patient1
- Paralysis1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6) 2019 A PATIENT UNDERWENT A POSTERIOR LUMBAR INTERBODY FUSION WITH A COMPETITOR'S CAGE AND NUVASIVE PRODUCTS AT L4/5. ON (B)(6) 2020 A REVISION SURGERY WAS PERFORMED DUE TO THE COMPETITOR'S CAGE BACKING OUT. IT WAS ALSO FOUND THAT THE NUVASIVE POSTERIOR FIXATION HAD LOOSENED, THEREFORE THE FIXATION AND COMPETITOR'S CAGE WAS REPLACED. NON-UNION WAS OBSERVED AS WELL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11158625 ↗MalfunctionJul 1, 2019Break; Break
ON (B)(6) 2019, PATIENT UNDERWENT A REVISION PROCEDURE TO REMOVE A PREVIOUSLY IMPLANTED CONSTRUCT AT THE L3 TO L4 VERTEBRAL LEVELS AND TO EXTEND THE CONSTRUCT FROM VERTEBRAL LEVELS L4 TO S1. THE REVISION WENT AS PLANNED WITH NO ISSUES. IN A SUBSEQUENT ROUTINE 4 MONTH FOLLOW UP, IT WAS DISCOVERED THAT THERE WAS A UNILATERAL SCREW BREAKAGE AT THE S1 LEVEL. THE PATIENT IS CURRENTLY ASYMPTOMATIC AND NO REVISION PROCEDURE IS CURRENTLY PLANNED.
Patient problems: Failure of Implant
MAUDE report 8750431 ↗MalfunctionNov 20, 2018Break; Break
ON (B)(6) 2018, PATIENT UNDERWENT AN EXTREME LATERAL INTERBODY FUSION PROCEDURE AT T10-T11 LEVELS AND A POSTERIOR SPINAL FUSION PROCEDURE AT T10-PELVIC LEVELS. A POST-OPERATIVE CT IMAGE REVEALED A POSSIBILITY OF SCREW FRACTURE AT T10 AND T11.
Patient problems: Failure of Implant
MAUDE report 8090051 ↗MalfunctionMar 27, 2018Unintended Movement; Unintended Movement
ON (B)(6) 2018 PATIENT UNDERWENT A POSTERIOR LUMBAR FUSION PROCEDURE AT L4/5 LEVEL. THE POSITION OF THE IMPLANTS WERE CONFIRMED BY X-RAY IMAGES AND THE SURGERY WAS COMPLETED. REPORTEDLY, TWO DAYS POST-OPERATIVE THE PATIENT'S LEFT LEG BECAME PARALYZED. A CT IMAGE CONFIRMED THAT THE L5 SCREW HAD PENETRATED THE INNER BONE. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2018 TO REPLACE THE LEFT L5 SCREW WITH A HOOK. PATIENT HAS BEEN REPORTED TO BE DOING WELL POST REVISION PROCEDURE.
Patient problems: Failure of Implant; Paralysis
MAUDE report 7374113 ↗MalfunctionAug 24, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
ON (B)(6) 2017, A PATIENT UNDERWENT POSTERIOR SPINAL FUSION AT THE L3-SAI LEVELS. DURING A FOLLOW-UP, AN X-RAY FILM INDICATED TWO SCREWS HAD LOOSENED AT THE L3 LEVEL. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2017, TO REMOVE THE LOOSENED SCREWS, ADDED TWO SCREWS AT THE L2 LEVEL AND REPLACE THE RODS. IT WAS ALSO REPORTED NESPLON CABLES WERE ADDED AT THE L2-L3 LEVELS. REPORTEDLY PATIENT IS DOING WELL POST-SURGERY.
Patient problems: Failure of Implant
MAUDE report 6819220 ↗
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.