ICU Medical CL-10-10Recalls & reported problems

Icu Medical, Inc. · ChemoLock™ Bag Spike, 10 units · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887709071025), received between Jul 29, 2020 and Jul 29, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Disconnection” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

58 other recalls cover devices cleared under the same FDA submission (K131549); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 8, 2025Class IIOpen, Classified

    Port weld of drug transfer device may separate or break during use and potentially result in a leak.

    Z-0328-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 27, 2022Class IIOpen, Classified

    Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

    Z-1681-2022 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Disconnection1
  • Fluid/Blood Leak1

Patient problems

  • Chemical Exposure1
  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    THE EVENT INVOLVED A CHEMOLOCK¿ BAG SPIKE, 10 UNITS THAT THE CUSTOMER REPORTED LEAKING. THE UNKNOWN HAZARDOUS DRUG LEAKING WAS NOTICED DURING PRIMING FOR 1 HOUR AT 9:00 AM. THE CUSTOMER STATED THAT NO DRUG WAS LOST, HOWEVER, FLUID DRIPPED OUT UPON DISCONNECTING TO DISPOSE OF THE BAG, EXPOSING THE HAZARDOUS DRUG. WHEN THE CHEMOLOCK INJECTOR WAS DISCONNECTED FROM THE CHEMOLOCK BAG SPIKE THE RESIDUAL IN THE BAG SPILLED OUT ONTO THE FLOOR, EXPOSING THE HAZARDOUS DRUG. THERE WAS NO PATIENT INVOLVEMENT, NO ADVERSE EVENT, AND NO DELAY IN THERAPY.

    Patient problems: Chemical Exposure; No Patient Involvement

    MAUDE report 10337927 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131549. One submission often covers many models, so check the product and lot details in each record.

  • Z-3161-2026 · May 12, 2026 · Open, Classified
    IV Sets with drip chambers
  • Z-3159-2026 · May 12, 2026 · Open, Classified
    Oncology sets with drip chambers
  • Z-2210-2026 · Mar 24, 2026 · Open, Classified
    Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL
  • Z-0341-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Bag Spike REF: CL-10
  • Z-0361-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical 6" (15 cm) Ext Set w/MicroClave" Clear, ChemoLock" Port, Y-Connector, Rotating Luer REF: CL4118
  • Z-0356-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Syringe Transfer Set w/MicroClave, ChemoLock Port REF: CL-34
  • Z-0345-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Vented Vial Spike, 13mm REF: CL-72
  • Z-0337-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Oncology Transfer Kit w/ChemoClave and ChemoLock Port; Spiros w/Red Cap; ChemoLock Port REF: CL4138
  • Z-0331-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Clave Bag Spike w/ChemoLock Port REF: CL2150
  • Z-0360-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock" Port, Clamp, Graduated Adapter REF: CL3967
  • Z-0365-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemosafeLock Vial Adapter REFs: KL-VA001U3 KL-VA002U3 KL-VA131U3 KL-VA201U3 KL-VA202U3 KL-VA321U3
  • Z-0329-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Chemo Lock Universal Vented Vial Spike REF: CL-70-10
  • Z-0346-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Closed Vial Spike REF: CL-80
  • Z-0338-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer REF: CL4151
  • Z-0336-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Oncology Kit w/ChemoLock, ChemoLock Port REF: CL4136

…and 43 more.

Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices

All Closed antineoplastic and hazardous drug reconstitution and transfer system devices →