Skip to content

Product Description/Item: ChemoLock Bag Spike, 10… recallZ-1681-2022 · ICU Medical, Inc. · initiated Jun 27, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

FDA root cause category: Component design/selection

What the company is doing

On 06/27/2022 recall notices were mailed to customers who were asked to do the following: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) In instances where the benefits of using the ChemoLock are greater than the potential risks for non-connection or disconnection, and you choose to utilize ChemoLock affected lot numbers, ensure full engagement between the ChemoLock Port and ChemoLock Injector. If full engagement is achieved, the device will work as intended. If full engagement cannot be achieved or is difficult to achieve, discard product and utilize a new device. 3) Complete and return the response form. 4) The recalling firm's customer service can be contacted for product availability information. 5) Return affected product using the return label provided. Contact Sedgwick if you have not received a return label or require additional labels. 6) If you have distributed the product further, immediately notify your accounts and ask them to contact Sedgwick. For further inquiries, please contact: Global Complaint Management 1-844-654-7780 or [email protected]; Customer Service 1-866-829-9025, option 8 or [email protected]; Sedgwick 1-888-943-5190 (M-F, 8am-5pm ET)

Recall details

Recall number
Z-1681-2022
Status
Open, Classified
Initiated by the company
Jun 27, 2022
Recalling firm
ICU Medical, Inc.
Quantity
454,635
Product code
ONB
FDA submissions
K131549
Distribution
Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

Product description

Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Item/UDI-DI/Lots: CL-10-10/10887709071022/5872839, 5891255; CL-12/10887709056975/5819981, 5856055, 5873186; CL-13/10887709056982/5850811; CL2100/5797867, 5816334, 5829882,5850510; CL2100-5/10887709122960/5754377; CL2100T/10887709099293/5816341, 5842229, 5850517; CL2150/10887709094670/5829900, 5850507; CL3011/10887709057019/5829885, 5842227; CL3364/10887709073408/5841261, 5850805; CL-34/10887709069777/5754437, 5829875; CL3511/10887709057026/5842231, 5861991, 5894477; CL3511T/10887709120010/5829831, 5891263; CL3528/10887709057040/5850500, 5872833, 5883793, 5894481, 5909807; CL3535T/10887709099613/5685350; CL3538/10887709071909/5935851; CL3900/10887709071411/5772697; CL3927/10887709075440/5842211; CL3946/10887709080291/5829895, 5861983; CL3947/10887709080307/5782844; CL3952/10887709081571/5782863,5816379; CL3955/10887709082509/5841209, 5850839; CL3960/10887709083278/5816328, 5842216; CL3963/10887709083827/5842214,5851775; CL4114/10887709091136/5754773, 5782864, 5816356; CL4130/10887709095530/5850515; CL4131/10887709095776/5895940; CL4136/10887709096131/5816325, 5850661; CL4143/10887709097213/5782848, 5873243, 5891266; CL4146-5/10887709097466/5872843, 5883800, 5909812; CL4153/10887709098722/5816348; CL4159/10887709121369/5881696; CL4179/10887709123578/5872687; CL-62/10887709057088/5816320,5825248; CL-70/10887709057095/5842228; CL-70-10/10887709070988/5816343, 5851774; CL-70-10T/10887709099477/5816353; CL-70S/10887709067155/5842225,5850512; CL-70S-5/10887709070995/5816326, 5851776; CL-70T/10887709099507/5816350; CL-72/10887709062440/5829832, 5849740, 5816312; CL-72T/10887709099552/5841181, 5849748; CL-76/10887709122878/5816323, 5842218; CL-80/10887709057125/5804034, 5842200, 5850499; CL-80S/10887709069593/5808063, 5816306, 5829020, 5829023, 5841982, 5845354, 5862839; CL-80S-10/10840619080577/5804045, 5816315; CL-80S-10T/10887709120539/5804051, 5816310, 5829022, 5841994, 5849630, 5850497, 5871336; CL-80S-5/10887709077116/5667054, 5829021, 5829024, 5841980, 5849474, 5849631, 5849632, 5849713, 5849714, 5849715, 5849716, 5849474; CL-80SL/10887709069630/5816317; CL-82/10887709085692/5804065; CLH-12/10887709080413/5842212

Devices named in this recall

49 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

42 devices cleared under K131549 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates