ICU Medical CL3538Recalls & reported problems
Icu Medical, Inc. · Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock™ Additive Port, Vented Cap; ChemoLock™; Spinning Spiros® w/Red Cap · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887709071902), received between Jul 8, 2020 and Mar 5, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
59 other recalls cover devices cleared under the same FDA submission (K131549); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 27, 2022Class IIOpen, Classified
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
Z-1681-2022 · Root cause: Component design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Chemical Exposure1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2026Disconnection; Fluid/Blood Leak
THE EVENT INVOLVED AN ONCOLOGY KIT ADD-ON SET W/ CHEMOLOCK ADDITIVE PORT, VENTED CAP CHEMOLOCK SPINNING SPIROS WHERE THE CUSTOMER REPORTED THAT THE SPIROS DETACHED FROM A PRIMARY LINE DURING DRUG ADMINISTRATION (TAXOL). THE REPORT ALSO STATED THAT A MINIMAL SPILL AND LEAK WAS NOTICED IMMEDIATELY. THERE WAS NO REPORT OF PATIENT HARM AS A CONSEQUENCE OF THIS EVENT.
Patient problems: Chemical Exposure; No Clinical Signs, Symptoms or Conditions
MAUDE report 24522939 ↗MalfunctionJun 28, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED AN ONCOLOGY KIT W/5" (13 CM) ADD-ON SET W/CHEMOLOCK¿ ADDITIVE PORT, VENTED CAP; CHEMOLOCK¿; SPINNING SPIROS® W/RED CAP. THE CUSTOMER STATED THAT THE LEAK SEEMED TO BE COMING FROM THE JUNCTURE OF THE SPIROS AND LINE; IT IS UNCLEAR IF THEY ARE COMING FROM THE DISTAL END OF THE SPIROS THAT CONNECTS TO THE PATIENT, OR THE PROXIMAL END THAT CONNECTS TO THE PRIMARY TUBING, BUT IT IS BELIEVED TO BE PROXIMAL END THAT CONNECTS TO THE PRIMARY TUBING. THERE WAS PATIENT INVOLVEMENT BUT HARM WAS NOT REPORTED AS A CONSEQUENCE OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19634924 ↗MalfunctionJul 8, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A ONCOLOGY KIT W/5" (13 CM) ADD-ON SET W/CHEMOLOCK¿ ADDITIVE PORT, VENTED CAP; CHEMOLOCK¿; SPINNING SPIROS® W/RED CAP THAT THE CUSTOMER REPORTED AN UNSPECIFIED CHEMOTHERAPY MEDICATION LEAKING SPECIFICALLY WITH FROM THE CHEMOLOCK BAG SPIKE COMPONENT OF THE KIT OUT FROM A MATING DEVICE. THE LEAKING OCCURRED IMMEDIATELY AFTER THE DRUG WAS ADDED TO THE BAG. THE TUBING SET UP WAS THE 5" ADD-ON-SET SPIKED INTO BAG, UNSPECIFIED TUBING SET CONNECTED TO ADD-ON-SET, AND CH-2000SC ATTACHED TO THE END OF THE TUBING. THE CHEMOTHERAPY SPILL WAS CLEANED UP PER PROTOCOL AND THERE WAS NO SPILL KIT USED. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENT.
Patient problems: No Patient Involvement
MAUDE report 10244145 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131549. One submission often covers many models, so check the product and lot details in each record.
- Z-3159-2026 · May 12, 2026 · Open, ClassifiedOncology sets with drip chambers
- Z-3161-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-2210-2026 · Mar 24, 2026 · Open, ClassifiedOncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL
- Z-0358-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical 8" (20 cm) Ext Set w/MicroClave Clear, ChemoLock Port, Y-Connector, Rotating Luer, REF: CL-3950
- Z-0355-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock w/Mini Bag Spike REF: C-17
- Z-0350-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Port REF: CL2100
- Z-0363-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemosafeLock Bag Spike REF: KL-BS001U3
- Z-0330-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock" Port, Bulk, Non-Sterile REF: CL2100-NS
- Z-0339-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical 60" (152 cm) Appx 1.2 ml, Ext Set, Smallbore w/ChemoLock Port, Clamp, Spiros w/Red Cap REF: CL4173
- Z-0362-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Oncology Kit w/2 ChemoLock", 13mm Closed Vial Spike w/ChemoLock" Port, ChemoLock" Bag Spike, Ext Set w/ChemoLock" Port, Clamp, Graduated Adapter REF: CL4188
- Z-0336-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Oncology Kit w/ChemoLock, ChemoLock Port REF: CL4136
- Z-0346-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Closed Vial Spike REF: CL-80
- Z-0364-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemosafeLock Connecter REF: KL-FNU3
- Z-0349-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock 13mm Closed Vial Spike REF: CL-82
- Z-0357-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Port w/Spiros, Red Cap REF: CL-35
…and 44 more.
Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices
- BD PhaSeal 515003BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall183 reports
- BD PhaSeal 515056BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall159 reports
- BD PhaSeal Optima 515070BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall154 reports
- Alaris 10012241-0500BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall138 reports
- BD PhaSeal 515005BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall128 reports
- ICU Medical CL3011Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls26 reports
- BD PhaSeal Optima 515052BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall114 reports
- ICU Medical CL3511Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page