Skip to content

ICU Medical CL2100TRecalls & reported problems

Icu Medical, Inc. · ChemoLock™ Port · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887709099296), received between Oct 29, 2023 and Oct 29, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Free or Unrestricted Flow” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

59 other recalls cover devices cleared under the same FDA submission (K131549); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 27, 2022Class IIOpen, Classified

    Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

    Z-1681-2022 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Free or Unrestricted Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2023Free or Unrestricted Flow; Free or Unrestricted Flow

    THE EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A CHEMOLOCK¿ PORT WHERE IT WAS REPORTED THAT THE PRODUCT IS PULLING FROM THE PRIMARY BAG AND SECONDARY BAG AT THE SAME TIME. THE EVENT OCCURRED ON DURING INFUSION. THE PRODUCT HAS BEEN USED BEFORE THE ISSUE WAS NOTICED AND THERE WAS NO LEAKING. THERE WAS PATIENT INVOLVED, AND UNKNOWN PATIENT HARM, BUT THE ISSUE RESULTED INTO PROLONGED INFUSION TIME AND THERE WAS DELAY IN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18026873 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131549. One submission often covers many models, so check the product and lot details in each record.

  • Z-3159-2026 · May 12, 2026 · Open, Classified
    Oncology sets with drip chambers
  • Z-3161-2026 · May 12, 2026 · Open, Classified
    IV Sets with drip chambers
  • Z-2210-2026 · Mar 24, 2026 · Open, Classified
    Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL
  • Z-0345-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Vented Vial Spike, 13mm REF: CL-72
  • Z-0352-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Graduated Connector w/ChemoLock Port REF: CL3900
  • Z-0350-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Port REF: CL2100
  • Z-0336-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Oncology Kit w/ChemoLock, ChemoLock Port REF: CL4136
  • Z-0346-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Closed Vial Spike REF: CL-80
  • Z-0343-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Universal Vented Vial Spike REF: CL-70
  • Z-0342-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Double ChemoLock Port Syringe Transfer Set REF: CL-33
  • Z-0353-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical 6.5" Bag Spike w/ChemoLock Port Additive Port, ChemoLock Port REF: CL3955
  • Z-0340-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical 73" (185 cm) Appx 2.5 ml, Ext Set, Smallbore w/ChemoLock Port, Clamp, ChemoLock w/Red Cap REF: CL4176
  • Z-0364-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemosafeLock Connecter REF: KL-FNU3
  • Z-0344-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical ChemoLock Vented Vial Spike, 20mm REF: CL-70S
  • Z-0334-2026 · Sep 8, 2025 · Open, Classified
    ICU Medical Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock" Port Drip Chamber, Spiros"; ChemoLock"; ChemoLock" Universal Vented Vial Spike; Bag Hanger REF: C

…and 44 more.

Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices

All Closed antineoplastic and hazardous drug reconstitution and transfer system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page