iLED7 Surgical Light 4068110Recalls & reported problems
Baxter Medical Systems GmbH + Co. KG · iLED 7 Ceiling Single · Class 2
All models of iLED7 Surgical Light (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761968325), received between Jun 12, 2025 and Feb 12, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Unintended Movement” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jun 4, 2024Open, Classified
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
Z-2315-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Unintended Movement1
Patient problems
- Bruise/Contusion2
- Hematoma2
- Concussion1
- Head Injury1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem; Unintended Movement
IT WAS REPORTED A SURGEON HIT THEIR HEAD ON AN ILED LIGHT AND SUSTAINED A CONCUSSION. THE SURGEON WAS PERFORMING A PROCEDURE ON A PATIENT WHEN THE PHYSICIAN ¿MOVED TO GET AROUND THE PATIENT'S KNEE¿ AND HIT THEIR HEAD ¿VIOLENTLY¿ ON THE HANDLE/CAMERA OF THE LIGHT FIXTURE. THE SURGEON COMPLAINED OF PAIN AND DIZZINESS. THE SURGEON HAD TO SIT DOWN ¿UNTIL THEY REGAINED THEIR SENSES.¿ THE SURGEON EXPERIENCED PAIN FOR A FEW DAYS IN WHICH THEY TOOK UNSPECIFIED ¿PAIN KILLERS.¿ CURRENTLY THE PHYSICIAN HAS RECOVERED. ADDITIONALLY, IT WAS REPORTED THE SURGEON WAS TALL AND THE OPERATING ROOM ¿THE ILED7 LIGHTS REMAIN TO BE TOO LOW DESPITE THE CHANGE OF THE LIGHTS SPRING ARM FROM NORMAL HEIGHT TO THE ONES DEDICATED FOR LOW CEILINGS¿ IN 2025. THE ILED 7DEVICE INSTRUCTION FOR USE WARNS ¿A COLLISION INVOLVING THE SUPPORT ARMS OR LAMP HEAD CAN RESULT IN DAMAGE. AVOID COLLISIONS WITH OTHER OBJECTS OR ADJACENT WALLS. BEFORE ADJUSTING THE HEIGHT, CHECK THAT THERE IS SUFFICIENT CEILING CLEARANCE AND ENSURE THAT THERE ARE NO OTHER OBJECTS POSITIONED ABOVE THE LAMP HEAD. FURTHERMORE, THE INSTRUCTIONS FOR USE STATE ¿BEFORE THE MEDICAL DEVICE IS APPLIED, THE USER MUST ENSURE THAT THE PRODUCT IS OPERATIONAL A
Patient problems: Concussion; Head Injury; Pain
MAUDE report 24329435 ↗InjuryJun 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AN UNSPECIFIED QUANTITY OF HEALTHCARE PROFESSIONALS SUSTAINED AN ARRAY OF ¿HEAD INJURIES¿ DUE TO THE HEIGHT OF THE ILED7 SURGICAL LIGHT. IN OPERATING ROOM #6 THE CEILING IS LOW, AND THIS CAUSED THE HEIGHT OF THE ILED SURGICAL LIGHT TO BE LOWER AS WELL. FURTHERMORE, THE NURSES AND SURGEONS HAVE SUSTAINED CONTUSIONS AND HEMATOMAS. ONE NURSE IS ON ¿SICK LEAVE¿ AFTER SUSTAINING AN UNSPECIFIED ¿HEAD INJURY¿ FROM THE IMPACT WITH THE OPERATING LIGHT. THERE WAS NO ALLEGED MALFUNCTION OF THE ILED LIGHT. ADDITIONALLY, THE CUSTOMER CONTINUED TO USE THE ALLEGED ¿LOW LIGHTS¿ POST-INSTALLATION, THUS, USE ERROR CANNOT BE EXCLUDED AS A CONTRIBUTING FACTOR TO THE INJURY. FURTHERMORE, PREVENTION OF THIS TYPE OF EVENT IS OUTLINED IN THE DEVICE INSTRUCTIONS FOR USE. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Bruise/Contusion; Hematoma
MAUDE report 22194053 ↗InjuryJun 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AN UNSPECIFIED QUANTITY OF HEALTHCARE PROFESSIONALS SUSTAINED AN ARRAY OF ¿HEAD INJURIES¿ DUE TO THE HEIGHT OF THE ILED7 SURGICAL LIGHT. IN OPERATING ROOM # 6 THE CEILING IS LOW, AND THIS CAUSED THE HEIGHT OF THE ILED SURGICAL LIGHT TO BE LOWER AS WELL. FURTHERMORE, THE NURSES AND SURGEONS HAVE SUSTAINED CONTUSIONS AND HEMATOMAS. ONE NURSE IS ON ¿SICK LEAVE¿ AFTER SUSTAINING AN UNSPECIFIED ¿HEAD INJURY¿ FROM THE IMPACT WITH THE OPERATING LIGHT. THERE WAS NO ALLEGED MALFUNCTION OF THE ILED LIGHT. ADDITIONALLY, THE CUSTOMER CONTINUED TO USE THE ALLEGED ¿LOW LIGHTS¿ POST-INSTALLATION, THUS, USE ERROR CANNOT BE EXCLUDED AS A CONTRIBUTING FACTOR TO THE INJURY. FURTHERMORE, PREVENTION OF THIS TYPE OF EVENT IS OUTLINED IN THE DEVICE INSTRUCTIONS FOR USE. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Bruise/Contusion; Hematoma
MAUDE report 22194044 ↗
Other Lamp, operating-room devices
- TL 5000/3000 – Surgical Light 4038310Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room1 recall1 report
- iLED7 Surgical Light 4068410Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room1 recall0 reports
- iLED 4068210Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room12 reports
- SynFrame 387.345Synthes GmbH · Lamp, operating-room12 reports
- SynFrame 387.338Synthes GmbH · Lamp, operating-room9 reports
- SynFrame 387.343Synthes GmbH · Lamp, operating-room9 reports
- iLED 4068310Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room8 reports
- SynFrame 387.336Synthes GmbH · Lamp, operating-room6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page