Baxter Medical Systems GmbH + Co. KG · Hospital Device · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00887761995840), received between Sep 29, 2022 and Mar 30, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Excessive Heating2
- Insufficient Device Problem Information2
- Mechanical Problem2
- Improper or Incorrect Procedure or Method1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Detachment of Device or Device Component
BAXTER RECEIVED A REPORT THAT ILED7 TRIO/QUAD HAD SPRING ARM FALLING OFF THE CENTER AXIS. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24736185 ↗MalfunctionFeb 28, 2025Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information
THE CUSTOMER ALLEGED THAT DURING A SURGICAL PROCEDURE, THE ILED SURGICAL LIGHT SYSTEM BURNED A PATIENT. FOLLOW-UP WITH THE CUSTOMER FOUND THAT THE BURN WAS REPORTED TO BE A FULL THICKNESS BURN AND ¿WOULD REQUIRE ADDITIONAL SURGERY TO MANAGE THE BURN. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 21490523 ↗MalfunctionApr 5, 2024Excessive Heating; Excessive Heating
IT WAS REPORTED THAT WHEN THE FACILITY MANAGER EXPERIMENTED THE ILED 7 CEILING TRIO LIGHTS ON HIMSELF HE FELT WARM UNDER THE OT SCRUB AND OBSERVED REDNESS AFTER 20 MINS. IT WAS NOTED THAT THE 3 LIGHTS WERE BEING USED AT THE SAME TIME, AND SET AT 100% INTENSITY AND 5000K COLOUR TEMPERATURE. NO MEDICAL INTERVENTION WAS REQUIRED. DURING FOLLOW-UP WITH THE CUSTOMER, IT WAS NOTED THAT THE CUSTOMER WAS UNAWARE OF THE REMAINING RISK ASSOCIATED WITH THE OVERLAPPING OF 2 OR MORE LIGHTS. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19051040 ↗MalfunctionApr 5, 2024Insufficient Device Problem Information; Excessive Heating; Insufficient Device Problem Information; Excessive Heating
IT WAS REPORTED THAT DURING AN OT CASE SIMULATION WITH THE ILED 7 CEILING TRIO LIGHTS, ONE OF THE NURSE WAS ON THE OPERATING TABLE. THE LIGHT WAS SET AT 100% INTENSITY, COLOUR TEMPERATURE 5000K, SHINING ON THE NURSED KNEE, WHO WAS WEARING OT SCRUB. AFTER APPROXIMATELY 6 TO 10 MINUTES, THE NURSE FELT DISCOMFORT ON THE KNEE, AND WHEN THEY ROLLED UP THE SCRUB, THEY FOUND THERE WAS REDNESS ON THE KNEE. NO MEDICAL INTERVENTION WAS REQUIRED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19050492 ↗MalfunctionDec 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE LIGHT HEAD OF THE ILED 7 SURGICAL LIGHT SYSTEM FELL DOWN DURING A SURGICAL PROCEDURE ON THE BACK OF A SURGEON. NO HARM OR IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18364275 ↗MalfunctionOct 13, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER REPORTED DURING A CASE, A SCREW FELL OUT OF THE MONITOR HOLDER AND LANDED ON TO THE FLOOR AFTER 3RD PARTY INSTALLED THE MONITORS ON THE CUSTOMERS SYSTEM. NO HARM OR IMPACT TO A SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15594658 ↗MalfunctionSep 30, 2022Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
THE CUSTOMER ALLEGED THE LIGHT HEAD WAS DROPPED. THE BIO MED WENT TO SECURE THE LIGHT HEAD TO THE SPRING ARM AND IT FELL TO THE FLOOR. NO HARM OR IMPACT TO A SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15518513 ↗MalfunctionSep 29, 2022Mechanical Problem; Detachment of Device or Device Component; No Apparent Adverse Event; Mechanical Problem; Detachment of Device or Device Component; No Apparent Adverse Event
THE CUSTOMER REPORTED THE PLATE WITH VESA MONITOR ADAPTER AND HANDPIECE HOLDER FELL DURING CLINICAL ACTIVITY WHICH IS PART OF THE MONITOR HOLDER ADAPTER FROM THE COMPANY CIM. NO HARM OR IMPACT TO A SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15507924 ↗
Other Lamp, operating-room devices
- iLED7 Surgical Light 4068110Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room1 recall3 reports
- TL 5000/3000 – Surgical Light 4038310Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room1 recall1 report
- iLED7 Surgical Light 4068410Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room1 recall0 reports
- SynFrame 387.345Synthes GmbH · Lamp, operating-room12 reports
- iLED 4068210Baxter Medical Systems GmbH + Co. KG · Lamp, operating-room12 reports
- SynFrame 387.343Synthes GmbH · Lamp, operating-room9 reports
- SynFrame 387.338Synthes GmbH · Lamp, operating-room9 reports
- SynFrame 387.336Synthes GmbH · Lamp, operating-room6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.