Skip to content

Likorall™ overhead lift 3122011Recalls & reported problems

Liko AB · Overhead track patient lifting/transfer system · Class 1

All models of Likorall overhead lift (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761979444), received between Feb 5, 2025 and Jan 26, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Fell” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 14, 2026Open, Classified

    Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.

    Z-0151-2027 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Fell2
  • Mechanical Problem2
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Fall2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2026Device Fell

    ON 30-DEC-2025, A CUSTOMER CONTACTED TECHNICAL SERVICE TO REPORT THAT LIKORALL 242 S R2R, NATURAL (PRODUCT CODE P3122011, SERIAL NUMBER (B)(6)), HAD PATIENT LIFTED OUT OF THE POOL WHEN THE DEVICE MALFUNCTIONED, CAUSING THE PATIENT TO FALL ON THE FLOOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: Fall

    MAUDE report 24179069 ↗
  • MalfunctionNov 6, 2025Insufficient Device Problem Information; Device Fell; Insufficient Device Problem Information; Device Fell

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE PATIENT DROPPED DURING LIFT FROM WHEELCHAIR,. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: Fall; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23484268 ↗
  • MalfunctionFeb 11, 2025Mechanical Problem; Mechanical Problem

    DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT LIKORALL 242 S R2R, HAD AN ISSUE, MISSING EMERGENCY STOP CORD. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21354737 ↗
  • MalfunctionFeb 5, 2025Mechanical Problem; Mechanical Problem

    DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT LIKORALL 242 S R2R, NATURAL WAS MISSING THE EMERGENCY STOP CORD. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21311014 ↗

Other Lift, patient, non-ac-powered devices

All Lift, patient, non-ac-powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page