Likorall™ overhead lift 3122007Recalls & reported problems
Liko AB · Overhead track patient lifting/transfer system · Class 1
All models of Likorall overhead lift (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761979758), received between Nov 6, 2024 and Mar 4, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 14, 2026Open, Classified
Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Z-0151-2027 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2025Mechanical Problem; Mechanical Problem
DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT LIKORALL 242 ES R2R, NATURAL HAD AN ISSUE, WHILE PERFORMING FUNCTIONS CHECK IT WAS NOTED THAT THE EMERGENCY STOP CORD WAS MISSING. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21513196 ↗MalfunctionNov 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 ES R2R OVERHEAD LIFT FELL OFF OF THE RAIL ONTO THE FLOOR, IT WAS ADDITIONALLY REPORTED THAT THE LIFT HAD A TURN TABLE ON IT GOING INTO TWO DIFFERENT BATHROOMS AND THE ACCOUNT TOOK IT UPON THEMSELVES TO TAKE DOWN THE RAILING GOING INTO ONE OF THE BATH ROOMS NOT REALIZING THAT THERE WAS NO WAY TO STOP THE LIFT FROM FALLING OFF ONTO THE FLOOR. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20734550 ↗MalfunctionNov 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE DISTRIBUTOR CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 ES R2R MOTOR FELL FROM RAIL. WHEN THE STAFF WERE MOVING THE HOIST BETWEEN TRACKS VIA THE TRACK GATE, THE HOIST MOVED OFF THE END OF THE TRACK AND FELL ON TO THE FLOOR DAMAGING THE END COVER. THE DEVICE WAS NOT BEING USED BY A PATIENT AT THE TIME. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20621056 ↗
Other Lift, patient, non-ac-powered devices
- MEDLINE MDS88200DMedline Industries, Inc. · Lift, patient, non-ac-powered108 reports
- Medline MDS600ELBMedline Industries, Inc. · Lift, patient, non-ac-powered82 reports
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered79 reports
- MEDLINE MDS88200HLMedline Industries, Inc. · Lift, patient, non-ac-powered77 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall24 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall18 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record