Likorall™ overhead lift 3122009Recalls & reported problems

Liko AB · Overhead track patient lifting/transfer system · Class 1

All models of Likorall overhead lift (25)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761979796), received between Mar 18, 2025 and Sep 26, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Fell” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 14, 2026Class IIOpen, Classified

    Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.

    Z-0151-2027 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Fell2
  • Insufficient Device Problem Information2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Erythema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2025Insufficient Device Problem Information; Device Fell; Insufficient Device Problem Information; Device Fell

    THE CUSTOMER REPORTED THAT "THE STRAP WAS DAMAGED AS A RESULT OF A FALL" THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23152966 ↗
  • MalfunctionMar 18, 2025Device Fell; Device Fell

    A CUSTOMER REPORTED THE LIKORALL 242 OVERHEAD LIFT MOTOR FELL ON THE FEMALE CAREGIVER CAUSING REDNESS TO HER RIGHT WRIST. TREATMENT WAS REPORTED AS APPLICATION OF HEMOCLAR CREAM. THERE WAS NO PATIENT IN THE LIFT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: Erythema

    MAUDE report 21631417 ↗
  • MalfunctionMar 18, 2025Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information

    A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 S, NATURAL, HAD AN ISSUE, CUSTOMER STATED THE SLING BAR IS DISENGAGED FROM THE STRAP. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21629785 ↗

FDA clearance and classification

The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Likorall™ overhead lift 3122009

Has the Likorall™ overhead lift 3122009 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0151-2027 (Class II), began on Sep 14, 2026: Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Likorall™ overhead lift 3122009?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 18, 2025 and Sep 26, 2025. In those reports, the most common specific device problems are “Device Fell” (2) and “Detachment of Device or Device Component” (1); the most common patient problem is “Erythema” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Likorall™ overhead lift 3122009?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Likorall™ overhead lift 3122009 FDA approved?

Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Liko AB.

Who makes the Likorall™ overhead lift 3122009, and is it still sold?

The FDA's device database lists Liko AB as the company responsible for it (brand Likorall™ overhead lift, model 3122009). Its record shows it as in commercial distribution.