Skip to content

Viking™ L mobile lift 2040044Recalls & reported problems

Liko AB · Mobile patient lifting system, battery-powered · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761981867), received between Oct 22, 2024 and May 7, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 30, 2025Open, Classified

    The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

    Z-2031-2025 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fracture1
  • Material Frayed1
  • Noise, Audible1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Fracture

    BAXTER RECEIVED A REPORT THAT VIKING L HAD SLING BAR HOOK BROKEN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25117195 ↗
  • MalfunctionNov 5, 2025Sparking; Noise, Audible; Sparking; Noise, Audible

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE VIKING L SOMETIMES SPARKS WHEN THE ELECTRICAL CONNECTION BETWEEN THE BATTERY AND THE SOCKET IS PLUGGED IN OR DISCONNECTED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23481820 ↗
  • MalfunctionJul 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A USER REPORT WAS RECEIVED BY BAXTER REPORTING THE FOLLOWING INCIDENT INVOLVING A VIKING L PATIENT LIFT. DURING A HOIST TRANSFER INVOLVING TWO PHYSIOTHERAPISTS AND ONE OCCUPATIONAL THERAPIST, A PATIENT WAS BEING TRANSFERRED FROM BED TO CHAIR USING A HOIST THAT WAS FULLY FUNCTIONAL AND WITHIN INSPECTION DATE. THE SLING WAS CORRECTLY POSITIONED AND ATTACHED PER STANDARD PROCEDURE. WHILE ELEVATING THE PATIENT, ONE OF THE CLIPS SECURING THE RIGHT LEG SLING UNEXPECTEDLY DETACHED, RESULTING IN THE LEG STRAP COMING OUT OF PLACE. THE PATIENT REMAINED FULLY SUPPORTED BY THE REMAINING THREE HOIST STRAPS. STAFF PROMPTLY AND SAFELY LOWERED THE PATIENT TO THE FLOOR, ENSURING SUPPORT OF THEIR HEAD AND LEGS THROUGHOUT. A HOVERJACK WAS USED TO TRANSFER THE PATIENT FROM THE FLOOR. A MEDICAL. REVIEW POST-INCIDENT CONFIRMED THAT THE PATIENT SUSTAINED NO KNOWN INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22608235 ↗
  • MalfunctionOct 22, 2024Material Frayed; Material Frayed

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE VIKING L PATIENT LIFT POWER CORD WAS FRAYING, STATED COPPER WIRES SHOWING. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509478 ↗

Other Lift, patient, non-ac-powered devices

All Lift, patient, non-ac-powered devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.