Viking™ L mobile lift 2040044Recalls & reported problems
Liko AB · Mobile patient lifting system, battery-powered · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761981867), received between Oct 22, 2024 and May 7, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 30, 2025Class IOpen, Classified
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
Z-2031-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Fracture1
- Material Frayed1
- Noise, Audible1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Fracture
BAXTER RECEIVED A REPORT THAT VIKING L HAD SLING BAR HOOK BROKEN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25117195 ↗MalfunctionNov 5, 2025Sparking; Noise, Audible; Sparking; Noise, Audible
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE VIKING L SOMETIMES SPARKS WHEN THE ELECTRICAL CONNECTION BETWEEN THE BATTERY AND THE SOCKET IS PLUGGED IN OR DISCONNECTED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23481820 ↗MalfunctionJul 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
A USER REPORT WAS RECEIVED BY BAXTER REPORTING THE FOLLOWING INCIDENT INVOLVING A VIKING L PATIENT LIFT. DURING A HOIST TRANSFER INVOLVING TWO PHYSIOTHERAPISTS AND ONE OCCUPATIONAL THERAPIST, A PATIENT WAS BEING TRANSFERRED FROM BED TO CHAIR USING A HOIST THAT WAS FULLY FUNCTIONAL AND WITHIN INSPECTION DATE. THE SLING WAS CORRECTLY POSITIONED AND ATTACHED PER STANDARD PROCEDURE. WHILE ELEVATING THE PATIENT, ONE OF THE CLIPS SECURING THE RIGHT LEG SLING UNEXPECTEDLY DETACHED, RESULTING IN THE LEG STRAP COMING OUT OF PLACE. THE PATIENT REMAINED FULLY SUPPORTED BY THE REMAINING THREE HOIST STRAPS. STAFF PROMPTLY AND SAFELY LOWERED THE PATIENT TO THE FLOOR, ENSURING SUPPORT OF THEIR HEAD AND LEGS THROUGHOUT. A HOVERJACK WAS USED TO TRANSFER THE PATIENT FROM THE FLOOR. A MEDICAL. REVIEW POST-INCIDENT CONFIRMED THAT THE PATIENT SUSTAINED NO KNOWN INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608235 ↗MalfunctionOct 22, 2024Material Frayed; Material Frayed
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE VIKING L PATIENT LIFT POWER CORD WAS FRAYING, STATED COPPER WIRES SHOWING. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20509478 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Viking™ L mobile lift 2040044
Has the Viking™ L mobile lift 2040044 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2031-2025 (Class I), began on May 30, 2025: The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Viking™ L mobile lift 2040044?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 22, 2024 and May 7, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1), “Fracture” (1) and “Material Frayed” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Viking™ L mobile lift 2040044?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Viking™ L mobile lift 2040044 FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Liko AB.
Who makes the Viking™ L mobile lift 2040044, and is it still sold?
The FDA's device database lists Liko AB as the company responsible for it (brand Viking™ L mobile lift, model 2040044). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, patient, non-ac-powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.