Skip to content

Multirall™ overhead lift 3130001Recalls & reported problems

Liko AB · Overhead track patient lifting/transfer system · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Liko Multirall overhead patient lift has 25 adverse event reports, with no deaths and 5 injuries. The reports mostly describe parts detaching or falling during use, straps tearing, jamming, or unspooling, and in a few cases patients falling or caregivers being struck. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00887761982284), received between Dec 7, 2023 and Jun 1, 2026 — classified as 4 injuries, 20 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component8
  • Mechanical Problem5
  • Device Fell3
  • Material Split, Cut or Torn3
  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Component Incompatible1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Laceration(s)2
  • Abrasion1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1
  • Erythema1
  • Head Injury1
  • Hematoma1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2026Failure to Auto Stop

    BAXTER RECEIVED A REPORT THAT MULTIRALL 200 HAD EMERGENCY STOP BUTTON INOPERATIVE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25356113 ↗
  • MalfunctionMay 15, 2026Detachment of Device or Device Component

    BAXTER RECEIVED A REPORT THAT MULTIRALL 200 HAD STRAP FALLING ON PATIENT. THIS OCCURRED DURING PATIENT USE. IT WAS REPORTED THAT THERE WAS NOT PATIENT/USER INJURY OR MEDICAL INTERVENTION WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25192445 ↗
  • MalfunctionMay 7, 2026Device Fell

    BAXTER RECEIVED A REPORT THAT MULTIRALL HAD ENGINE FALL. THE STRAP DETACHED FROM THE RAIL WITHOUT THE PATIENT PRESENT, MEDICAL STAFF INTERVENED TO SECURE THE MOTOR MID-AIR, BUT THE ATTACHMENT SUBSEQUENTLY FELL ONTO THE PATIENT. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS DURING USE. THERE WAS NOT PATIENT INJURY.

    Patient problems: Hematoma

    MAUDE report 25117340 ↗
  • MalfunctionApr 23, 2026Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT THAT MULTIRALL 200 HAD STRAP JAMMED INSIDE THE UNIT WHEN TRYING TO REWIND IT. WHEN PRESSING THE BUTTON THE STRAP FAILED TO RETRACT AND DISCOVERED IT WAS ALL SHREDDED. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24973581 ↗
  • InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    BAXTER RECIEVED A REPORT WHEN A CAREGIVER PUSHED THE LIFT TO REMOVE IT FROM THEIR WORKING AREA BUT PUSHED IT TOO HARD AND MADE IT SWING, AND WHEN IT CAME BACK, IT HIT THEM IN THE TEMPLE. THE CAREGIVER WAS ADMITTED TO THE EMERGENCY DEPARTMENT FOR A SUSPECTED HEAD INJURY AND WAS GIVEN A ONE-WEEK MEDICAL LEAVE. ADDITIONALLY, NO INITIAL DEVICE MALFUNCTION WAS REPORTED. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Head Injury

    MAUDE report 24665109 ↗
  • MalfunctionMar 13, 2026Detachment of Device or Device Component

    ON (B)(6) HAD A MOBILE LIFT MOTOR STRAP SUDDENLY UNSPOOLED, CAUSING A RESIDENT FALL. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: Abrasion; Erythema; Laceration(s)

    MAUDE report 24595660 ↗
  • MalfunctionJan 12, 2026Detachment of Device or Device Component

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT MULTIRALL 200 HAD AN ISSUE, DURING THE PATIENT'S TRANSFER FROM THE CHAIR TO THE BED, THE MOTOR DETACHED FROM THE RAIL. THE CAREGIVER WAS ABLE TO PUSH THE PATIENT BACK ONTO THE BED. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24038352 ↗
  • InjuryDec 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED A CAREGIVER SUSTAINED AN INJURY WHEN TRANSFERRING A PATIENT IN A MULTIRALL OVERHEAD LIFT. AFTER THE CAREGIVER HAD PROVIDED NURSING CARE TO A RESIDENT, THE CAREGIVER WANTED TO TRANSFER ANOTHER RESIDENT; HOWEVER, THE MOBILE PATIENT LIFT WASN'T WORKING BECAUSE THE BATTERY WASN'T PROPERLY SECURED. INSTEAD OF USING THE OTHER MOBILE PATIENT LIFT, THE CAREGIVER TOOK THE LIFT FROM THE ROOM ACROSS THE HALL, AND SINCE THE RAIL HEIGHTS WERE DIFFERENT, THE STRAP WAS SHORT, MAKING IT DIFFICULT TO ATTACH THE LIFTING ROD. THIS RESULTED IN THE DETACHMENT OF THE LIFT. THE CAREGIVER WAS ABLE TO SAVE THE RESIDENT WHO WAS ALREADY IN THE TRANSFER NET. THE PATIENT WAS NOT HARMED; HOWEVER, IT WAS NOTED THAT THE CAREGIVER REQUIRED SICK LEAVE DUE TO CERVICAL, SHOULDER, BACK, AND LEFT ARM INJURY. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23881582 ↗
  • MalfunctionNov 26, 2025Mechanical Problem; Key or Button Unresponsive/not Working; Mechanical Problem; Key or Button Unresponsive/not Working

    DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT MULTIRALL 200 (PRODUCT CODE 3130001, SERIAL NUMBER (B)(6), HAD EMERGENCY STOP BUTTON INOPERATIVE. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23662817 ↗
  • MalfunctionNov 5, 2025Mechanical Problem; Mechanical Problem

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT MULTIRALL 200 HAD AN ISSUE AND NEEDED REPAIR. DURING INSPECTION, THE BAXTER TECHNICIAN FOUND THE EMERGENCY STOP BUTTON WAS FAULTY. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23475953 ↗
  • MalfunctionSep 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    ANOTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE MULTIRALL 200 ENGINE COULD NOT BE CHARGED. PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23180844 ↗
  • MalfunctionAug 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE USE OF THE MULTIRALL 200, THE EXTENSION BELT CAME UNDONE, AND THE ENGINE FELL ON THE PATIENTS¿ ABDOMEN. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22837658 ↗
  • MalfunctionAug 1, 2025Mechanical Problem; Mechanical Problem

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT MULTIRALL 200 HAD AN ISSUE, BROKEN EMERGENCY STOP, RANDOM MALFUNCTION. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22687182 ↗
  • MalfunctionJul 2, 2025Mechanical Problem; Mechanical Problem

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE MULTIRALL 200 HAD AN ISSUE, UPON INSPECTION, THE BAXTER TECHNICIAN FOUND THE EMERGENCY STOP BUTTON WAS NOT FUNCTIONAL ON THE DEVICE. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22381946 ↗
  • MalfunctionApr 4, 2025Device Fell; Device Fell

    A USER REPORT WAS RECEIVED FROM THE MPA STATING THAT THE MULTIRALL 200 LIFT HAD AN ISSUE. WHEN MOUNTING THE LIFT UP, THE LIFT RELEASED THE BLACK RIBBON. THE LIFT FELL ONTO THE PATIENT'S BED. FORTUNATELY, THE PATIENT WAS LYING TO THE RIGHT AND THE LIFT ENDED UP ON THE LEFT SIDE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21771404 ↗
  • MalfunctionMar 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE MULTIRALL 200 LIFT FELL. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21707770 ↗
  • InjuryMar 19, 2025Disconnection; Disconnection

    IT WAS REPORTED A PATIENT SUSTAINED A LACERATION ABOVE THE RIGHT EYE WHEN USING A MULTIRALL OVERHEAD LIFT. DURING A PATIENT TRANSFER FROM THE BED TO A CHAIR, THE MULTIRALL 200 LIFT DETACHED FROM THE S65 HOOK, AND FELL ONTO THE PATIENT'S HEAD WHILE THEY WERE SITTING IN THE CHAIR. THE EVENT RESULTED IN A LACERATION ABOVE THE RIGHT EYE, WHICH RESULTED IN TWO STITCHES. THE PATIENT ALSO SUSTAINED A ¿POTENTIAL NASAL FRACTURE¿; HOWEVER, THERE IS NO FURTHER INFORMATION TO CONFIRM IF THERE WAS A FRACTURE. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Laceration(s)

    MAUDE report 21648586 ↗
  • MalfunctionFeb 18, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR CONTACTED TECHNICAL SERVICE TO REPORT THAT MULTIRALL 200 HAD AN ISSUE, THE LIFT STRAP TORE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21405301 ↗
  • MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE MULTIRALL 200 HAD AN ISSUE, THE EMERGENCY STOP SWITCH WAS FAULTY. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21110680 ↗
  • MalfunctionNov 19, 2024Component Incompatible; Component Incompatible

    A CUSTOMER CONTACTED TECHNICAL SERVICE TO REPORT THAT MULTIRALL 200 UNIT HAD A RECALL A FEW YEARS PRIOR AND THE CUSTOMER WAS NOT INFORMED BY THE 3RD PARTY THEY BOUGHT THE UNIT FROM NOR HAD THE CORRECTIVE ACTION PARTS BEEN REPLACED. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20726157 ↗
  • MalfunctionOct 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER ALLEGED THE LIFT¿S EMERGENCY BUTTON WAS OUT OF ORDER. THERE WAS NO ALLEGATION OF PATIENT OR CAREGIVER INJURY OR DEATH REPORTED FROM THIS ALLEGED INCIDENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514120 ↗
  • MalfunctionJul 16, 2024Break; Mechanical Problem; Break; Mechanical Problem

    THE CUSTOMER ALLEGED THE LIFT HAD BENT RODS ON THE INTERNAL MOTOR, CAUSING THE MOTOR TO TIGHTEN AND THE FUSE TO BE BLOWN. THERE WAS NO ALLEGATION OF PATIENT OR CAREGIVER INJURY OR DEATH REPORTED FROM THIS ALLEGED INCIDENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19748881 ↗
  • MalfunctionDec 19, 2023Break; Break

    THE CUSTOMER ALLEGED THE SLINGBAR WAS BROKEN NEAR THE SUPPORT HOOK AND NEEDED TO BE REPLACED. THERE WAS NO ALLEGATION OF PATIENT OR CAREGIVER INJURY OR DEATH REPORTED FROM THIS ALLEGED INCIDENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18354896 ↗
  • InjuryDec 7, 2023Device Fell; Device Fell

    THE CUSTOMER REPORTED THAT ON THE 1ST OF NOVEMBER 2023, A MULTIRALL LIFT FELL ON THE NURSE'S SHOULDER, AND THE NURSE WAS TAKEN TO THE EMERGENCY DEPARTMENT. ON THE 22ND OF NOVEMBER 2023, DURING FOLLOW-UP WITH THE HOSPITAL'S BIOMEDICAL, IT WAS REPORTED THAT THE NURSE SUFFERED OF A DOUBLE CLAVICLE (COLLARBONE) FRACTURE. SHE IS CURRENTLY IN SICK LEAVE AND WASN'T CURRENTLY ABLE TO PROVIDE INFORMATION ON THE MEDICAL TREATMENT SHE HAS / WILL RECEIVE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

    Patient problems: Bone Fracture(s)

    MAUDE report 18283158 ↗

FDA clearance and classification

The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Multirall™ overhead lift 3130001

Has the Multirall™ overhead lift 3130001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887761982284), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Multirall™ overhead lift 3130001?

The FDA's MAUDE database lists 24 adverse event reports for this device, received between Dec 7, 2023 and Jun 1, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (8), “Mechanical Problem” (5) and “Device Fell” (3); the most common patient problems are “Laceration(s)” (2), “Abrasion” (1) and “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Multirall™ overhead lift 3130001?

None of the reports are classified as deaths, 4 as injuries and 20 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Multirall™ overhead lift 3130001 FDA approved?

Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Liko AB.

Who makes the Multirall™ overhead lift 3130001, and is it still sold?

The FDA's device database lists Liko AB as the company responsible for it (brand Multirall™ overhead lift, model 3130001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Lift, patient, non-ac-powered devices

All Lift, patient, non-ac-powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page