Skip to content

Volara System M08085Recalls & reported problems

Hill-Rom Services Pte. Ltd. · Respiratory Care · Class 2

All models of Volara System (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 21, 2026Class IOpen, Classified

    There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

    Z-2549-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. May 21, 2026Class IOpen, Classified

    There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

    Z-2548-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code NHJ (device, positive pressure breathing, intermittent) under 21 CFR 868.5905 as a Class II device. 21 CFR 868.5905 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Volara System M08085

Has the Volara System M08085 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2549-2026 (Class I), began on May 21, 2026: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Volara System M08085 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Volara System M08085, and is it still sold?

The FDA's device database lists Hill-Rom Services Pte. Ltd. as the company responsible for it (brand Volara System, model M08085). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT devices

All DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT devices →