Volara System M08473Recalls & reported problems

Hill-Rom Services Pte. Ltd. · Respiratory Care · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 21, 2026Open, Classified

    There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

    Z-2550-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

  2. May 21, 2026Open, Classified

    There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

    Z-2551-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

  3. Apr 26, 2022Terminated

    There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.

    Z-1138-2022 · Root cause: Device Design · FDA record ↗

Other DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT devices

All DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT devices →