Percussionaire P5 (UDI 00849436000501)Recalls & reported problems
Percussionaire Corporation · IPV Phasitron Breathing Circuit Kit, Single Patient, Hospital Phasitron 5 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00849436000501), received between Oct 8, 2024 and Oct 9, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
4 FDA recall records name this device's identifier, 4 still open.
1 other recall cover devices cleared under the same FDA submission (K905236); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 18, 2026Class IIOpen, Classified
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Z-3322-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 18, 2026Class IIOpen, Classified
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Z-3321-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 18, 2026Class IIOpen, Classified
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Z-3324-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 18, 2026Class IIOpen, Classified
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Z-3323-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Component2
- Improper Chemical Reaction2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 9, 2024Defective Component; Improper Chemical Reaction; Defective Component; Improper Chemical Reaction
A COMPLAINT WAS RECEIVED ON (B)(6) 2024, FROM VA PALO ALTO MEDICAL CENTER REGARDING A BROWN LIQUID IN THE PHASITRON 5 TUBING. IT WAS STATED THAT THEY BELIEVE THE CIRCUIT WAS USED FOR 4 DAYS BEFORE NOTICING THE FLUID IN THE TUBING. NO ADVERSE REACTIONS HAVE BEEN REPORTED AT THIS TIME. THIS REPORT IS BEING COMPLETED RETROSPECTIVELY, AS IT HAS BEEN RECENTLY IDENTIFIED THAT THERE IS A POTENTIAL FOR HEALTH HAZARDS TO OCCUR DUE TO THIS ISSUE IF SPECIFIC PRODUCT LOTS ARE USED WITH CERTAIN MEDICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20408650 ↗MalfunctionOct 8, 2024Defective Component; Improper Chemical Reaction; Defective Component; Improper Chemical Reaction
A COMPLAINT WAS RECEIVED ON JULY 11, 2024, FROM TEXAS CHILDREN'S HOSPITAL REGARDING A BROWN LIQUID IN THE PHASITRON 5 TUBING. NO ADVERSE REACTIONS HAVE BEEN REPORTED AT THIS TIME. THIS REPORT IS BEING COMPLETED RETROSPECTIVELY, AS IT HAS BEEN RECENTLY IDENTIFIED THAT THERE IS A POTENTIAL FOR HEALTH HAZARDS TO OCCUR DUE TO THIS ISSUE IF SPECIFIC PRODUCT LOTS ARE USED WITH CERTAIN MEDICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20402144 ↗
FDA clearance and classification
Cleared through 510(k) K905236 on Apr 18, 1991 (decision: Substantially Equivalent). Submitted by Percussionaire Corp. as “Ipv Percussionator Family & Related Accessories”. The FDA review took 162 days. FDA record ↗
The FDA regulates product code NHJ (device, positive pressure breathing, intermittent) under 21 CFR 868.5905 as a Class II device. 21 CFR 868.5905 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K905236. One submission often covers many models, so check the product and lot details in each record.
- Z-0208-2025 · Sep 16, 2024 · Open, ClassifiedPhasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy
Common questions about the Percussionaire P5
Has the Percussionaire P5 been recalled?
Yes. 4 FDA recall records name this device's identifier: 4 still open. The most recent, Z-3322-2026 (Class II), began on Aug 18, 2026: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Percussionaire P5?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Oct 8, 2024 and Oct 9, 2024. In those reports, the most common specific device problems are “Defective Component” (2) and “Improper Chemical Reaction” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Percussionaire P5?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Percussionaire P5 FDA approved?
It was cleared through 510(k) K905236 on Apr 18, 1991. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Percussionaire P5, and is it still sold?
The FDA's device database lists Percussionaire Corporation as the company responsible for it (brand Percussionaire, model P5). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Positive Pressure Breathing, Intermittent devices
- Volara System M08473Hill-Rom Services Pte. Ltd. · DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT3 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls3 reports
- Volara System M07937Hill-Rom Services Pte. Ltd. · DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT2 recalls0 reports
- Volara System M08085Hill-Rom Services Pte. Ltd. · DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT2 recalls0 reports
- Percussionaire F00001-C-FRPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Percussionaire F00001-C-FRPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page