Percussionaire P5 (UDI 00849436000501)Recalls & reported problems

Percussionaire Corporation · IPV Phasitron Breathing Circuit Kit, Single Patient, Hospital Phasitron 5 · Class 2

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00849436000501), received between Oct 8, 2024 and Oct 9, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Defective Component” (2 reports).

4 FDA recall records name this device's identifier, 4 still open.

1 other recall cover devices cleared under the same FDA submission (K905236); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 18, 2026Open, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3324-2026 · Root cause: Device Design · FDA record ↗

  2. Aug 18, 2026Open, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3323-2026 · Root cause: Device Design · FDA record ↗

  3. Aug 18, 2026Open, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3322-2026 · Root cause: Device Design · FDA record ↗

  4. Aug 18, 2026Open, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3321-2026 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Defective Component2
  • Improper Chemical Reaction2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2024Defective Component; Improper Chemical Reaction; Defective Component; Improper Chemical Reaction

    A COMPLAINT WAS RECEIVED ON (B)(6) 2024, FROM VA PALO ALTO MEDICAL CENTER REGARDING A BROWN LIQUID IN THE PHASITRON 5 TUBING. IT WAS STATED THAT THEY BELIEVE THE CIRCUIT WAS USED FOR 4 DAYS BEFORE NOTICING THE FLUID IN THE TUBING. NO ADVERSE REACTIONS HAVE BEEN REPORTED AT THIS TIME. THIS REPORT IS BEING COMPLETED RETROSPECTIVELY, AS IT HAS BEEN RECENTLY IDENTIFIED THAT THERE IS A POTENTIAL FOR HEALTH HAZARDS TO OCCUR DUE TO THIS ISSUE IF SPECIFIC PRODUCT LOTS ARE USED WITH CERTAIN MEDICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20408650 ↗
  • MalfunctionOct 8, 2024Defective Component; Improper Chemical Reaction; Defective Component; Improper Chemical Reaction

    A COMPLAINT WAS RECEIVED ON JULY 11, 2024, FROM TEXAS CHILDREN'S HOSPITAL REGARDING A BROWN LIQUID IN THE PHASITRON 5 TUBING. NO ADVERSE REACTIONS HAVE BEEN REPORTED AT THIS TIME. THIS REPORT IS BEING COMPLETED RETROSPECTIVELY, AS IT HAS BEEN RECENTLY IDENTIFIED THAT THERE IS A POTENTIAL FOR HEALTH HAZARDS TO OCCUR DUE TO THIS ISSUE IF SPECIFIC PRODUCT LOTS ARE USED WITH CERTAIN MEDICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20402144 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905236. One submission often covers many models, so check the product and lot details in each record.

  • Z-0208-2025 · Sep 16, 2024 · Open, Classified
    Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Other Device, Positive Pressure Breathing, Intermittent devices

All Device, Positive Pressure Breathing, Intermittent devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.