Skip to content

NaviCare® Patient Safety Systems 3.9Recalls & reported problems

Hill-Rom, Inc. · Medical Software · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 19, 2023Class IIOpen, Classified

    Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.

    Z-2007-2023 · Root cause: Device Design · Recall details · FDA record ↗

Other MONITOR, BED PATIENT devices

All MONITOR, BED PATIENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page