POSEY 8345 (UDI 10190676002082)Recalls & reported problems
Posey Products, LLC · POSEY SITTER ELITE · Class 1
- Adverse event reports
- 12
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10190676002082), received between Sep 20, 2022 and Aug 23, 2024 — classified as 1 death, 11 malfunctions.
The most frequently reported specific device problem is “No Audible Alarm” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- No Audible Alarm9
- Communication or Transmission Problem1
- Fail-Safe Problem1
- Intermittent Loss of Power1
- Low Audible Alarm1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Bone Fracture(s)1
- Insufficient Information1
- Intracranial Hemorrhage1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2024No Audible Alarm; No Audible Alarm
UF/IMPORTER REPORT (B)(4). COMPLAINT REPORTED VIA THE FDA, SN (B)(6) ITEM # 8345, DISCONTINUED, ORDERED ON 5/17/22. COMPLAINT EVENT DATE WAS 5/17/24 EMAILED REPORT ON 7/26/24. IT WAS REPORTED A CHAIR ALARM DID NOT GO OFF AS DESIGNED. ALARM NOTED TO BE IN ON POSITION, BUT WHEN TESTED, DID NOT ACTIVATE WHEN SEAT BELT UNDONE, AS IT SHOULD. WHEN UNPLUGGED FROM SEAT BELT AND TURNED ON, ALARM SOUND WOULD WORK, BUT NOT WHEN PLUGGED IN. ALARM SWITCHED OUT, AND NEW ALARM WORKED WHEN ENGAGED AND PLUGGED IN AND SEAT BELT UNDONE, LEADING US TO SUSPECT PROBLEM WITH JACK PORT. ALARM PULLED FROM SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20054132 ↗MalfunctionJul 9, 2024No Audible Alarm; No Audible Alarm
SUPPLEMENTAL MEDWATCH BEING SENT FOR ADDITIONAL INFORMATION.
Patient problems: Bone Fracture(s)
MAUDE report 19702484 ↗MalfunctionMar 21, 2024No Audible Alarm; No Audible Alarm
TM REPORTED COMPLAINT VIA EMAIL. ISSUE OCCURRED WITH SN (B)(6) (EXPIRED ALARM), BUT RESULTED WITH A PATIENT INCIDENT/INJURY. PER THE TM: SITTER ELITE ALARM WITH THE 8399L-HEADSTART CHAIR BELT. ALARM DID NOT GO OFF WHEN PATIENT SELF-RELEASED THE HEADSTART BELT. PATIENT FELL AT 3 PM TODAY AND IS ON THEIR WAY TO HAVE CT SCAN TO IDENTIFY ANY INJURIES.
Patient problems: Insufficient Information
MAUDE report 18951378 ↗MalfunctionJan 16, 2024No Audible Alarm; Fail-Safe Problem; No Audible Alarm; Fail-Safe Problem
ALARM DOES NOT PROPERLY SOUND EVEN WITH ACTIVE POWER. MAGNET CORD WAS TESTED AND CONFIRMED THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18522188 ↗MalfunctionNov 16, 2023Intermittent Loss of Power; Intermittent Loss of Power
CUSTOMER CONTACTED VIA PHONE STATING THAT THE ALARM WILL CONTINUE TO ALARM INTERMITTENTLY. ISSUE ALSO OCCURS WHEN THE ALARM IS TURNED OFF. INITIAL TROUBLESHOOT VIA PHONE CORRECTED ISSUE, BUT CONTINUED ON WITHIN THE HOUR SINCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18154606 ↗MalfunctionNov 1, 2023Low Audible Alarm; Low Audible Alarm
CUSTOMER REPORTED COMPLAINT VIA PHONE CALL. PATIENT FALL OCCURRED WITH THE ALARM, VOLUME WAS QUIET UPON SOUNDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18052166 ↗MalfunctionMay 19, 2023No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED COMPLAINT VIA PHONE. ALARM WILL NOT POWER ON, TROUBLESHOOTING WITH NEW BATTERIES DID NOT RESOLVE ERROR, UNIT APPEARS TO HAVE NO DAMAGE. SN (B)(6). PLEASE ADVISE WHEN QA EVALUATION IS COMPLETE. CALLED IN BY (B)6). THE DATE THE ISSUE WAS DISCOVERED IS UNKNOWN AND NO INCIDENT OR SERIOUS INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16961392 ↗DeathOct 27, 2022No Audible Alarm; No Audible Alarm
TIDI PRODUCTS RECEIVED A LEGAL SUMMONS REGARDING A PATIENT DEATH AT SALEM REGIONAL MEDICAL CENTER WHILE UTILIZING POSEY SITTER ELITE ITEM #8345. SUMMONS NOTED INCORRECT SITTER ELITE MODEL NUMBER, HOWEVER THE MODEL NUMBER WAS CONFIRMED BY SALEM REGIONAL MEDICAL CENTER ATTORNEY. PATIENT WAS UNDER FALLS MANAGEMENT CARE PLAN, WHICH INCLUDED THE SITTER ELITE ALARM. A FALL WAS REPORTED THAT RESULTED IN AN INTRACRANIAL HEMORRHAGE THAT RESULTED IN A PATIENT DEATH. IN THE SUMMONS, IT INDICATES THAT THE ALARM DID NOT SOUND.
Patient problems: Intracranial Hemorrhage
MAUDE report 15679811 ↗MalfunctionSep 23, 2022Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2022 BD1 THE CUSTOMER CONTACTED US VIA PHONE CALL. CUSTOMER STATES THAT THE ALARM WILL NOT TRIGGER NURSE CALL SYSTEM AT THE NURSE STATION. CUSTOMER SAID THEY TIRED A NEW UNIT WITH WHERE THIS ALARM WAS AND IT WORKED. NO APPARENT DAMAGE TO THE UNIT ON THE NC PORT OR BATTERY COMPARTMENT. SN (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15470317 ↗MalfunctionSep 20, 2022No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED ONE ALARM VIA EMAIL. UNIT DOES NOT POWER ON, NO SIGNS OF CORROSION OR DAMAGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15460214 ↗MalfunctionSep 20, 2022No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED AN ALARM VIA EMAIL THAT THE ALARM DOES NOT MAKE SOUND. NO PHYSICAL DAMAGE OR MUTE MODE OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15457778 ↗MalfunctionSep 20, 2022No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED AN ALARM VIA EMAIL THAT THE ALARM DOES NOT MAKE SOUND. NO PHYSICAL DAMAGE OR MUTE MODE OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15457328 ↗
Other Monitor, Bed Patient devices
- POSEY 8645Posey Products, LLC · Monitor, Bed Patient1 death60 reports
- Voalte Patient Safety 4.0Hill-Rom, Inc. · MONITOR, BED PATIENT1 recall0 reports
- NaviCare® Patient Safety Systems 3.9Hill-Rom, Inc. · MONITOR, BED PATIENT1 recall0 reports
- Cubby Beds KIT10001Sensory Medical, Inc. · Monitor, Bed Patient30 reports
- Medline MDT8450YMedline Industries, Inc. · MONITOR, BED PATIENT29 reports
- POSEY 8399Posey Products, LLC · Monitor, Bed Patient19 reports
- POSEY 8309ELPosey Products, LLC · Monitor, Bed Patient2 deaths19 reports
- POSEY 8645WLPosey Products, LLC · Monitor, Bed Patient17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.