Skip to content

AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-380Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of AFFIXUSHIP FRACTURE NAIL - RIGHT (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887868039171), received between Jun 20, 2022 and Jun 20, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier.

249 other recalls cover devices cleared under the same FDA submission (K100238); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 7, 2017Class IITerminated

    During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.

    Z-0468-2018 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Bone Fracture(s)1
  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 20, 2022Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT HAD AN INITIAL SURGERY APPROXIMATELY 2 YEARS AGO, SHE HAS SINCE HAD A RIGHT BROKEN AFFIXUS HIP FRACTURE NAIL WHICH HAD TO BE REMOVED AND THE FRACTURE FIXED WITH A SYNTHES TROCHANTERIC HOOK PLATE. THE ORIGINAL FRACTURE WAS A SUB TROCHANTERIC FRACTURE EXTENDING DISTALLY TOWARDS THE MID SHAFT. A CABLE CAN BE SEEN ON THE X-RAY AT THE TIME OF REMOVING THE NAIL SHOWING SOME CALLOUS FORMATION AROUND THE MID TO PROXIMAL SECTION OF BONE AND LOOKS TO HAVE HEALED. THE NAIL HAD BROKEN AT THE CIRCLE SECTION WHERE THE LAG SCREW GOES THROUGH THE NAIL AND INTO THE FEMORAL NECK AND ACETABULA HEAD. THE FEMUR HAS FRACTURED FROM THE LATERAL SIDE APPROXIMATELY 8CM BELOW THE GREATER TROCHANTER AND EXTENDED MEDIALLY TOWARDS THE LESSER TROCHANTER AND INFERIOR FEMORAL NECK. THERE WAS NO DISCERNABLE REASON AS TO WHY THE NAIL HAD BROKEN. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Bone Fracture(s); Failure of Implant

    MAUDE report 14743667 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100238. One submission often covers many models, so check the product and lot details in each record.

  • Z-0447-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail Left Deg 9x300mm, Item Number 814609300 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedullar
  • Z-0448-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail Left 125 Deg 11x300mm, Item Number 814411300 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramed
  • Z-0450-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail Left 125 Deg 9x165mm, Item Number 816709165 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedu
  • Z-0451-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail 125 Deg 10x165mm, Item Number 816710165 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedullar
  • Z-0452-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail 125 Deg 12x165mm, Item Number 816712165 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedullar
  • Z-0449-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail Left 130 Deg 11x300mm, Item Number 814611300 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramed
  • Z-0453-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail 130 Deg 10x165mm, Item Number 816810165 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedullar
  • Z-0454-2019 · Oct 2, 2018 · Terminated
    Affixus Hip Fracture Nail 130 Deg 12x165mm, Item Number 816812165 The Affixus Hip Fracture Nail System, comprised of short and long nails, provides surgeons with an intramedullar
  • Z-0155-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail Left 125 13 mm x 340 mm, Item Number 814413340 Product Usage: Intended for the fixation of fractures
  • Z-0138-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail Right 125 13 mm x 180 mm, Item Number 814313180 Product Usage: Intended for the fixation of fractures
  • Z-0136-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail Right 125 11 mm x 440 mm, Item Number 814311440 Product Usage: Intended for the fixation of fractures
  • Z-0157-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail Right 130 9 mm x 320 mm, Item Number 814509320 Product Usage: Intended for the fixation of fractures
  • Z-0131-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail 125 11mm x 180mm, Item Number 814311180 Product Usage: Intended for the fixation of fractures
  • Z-0169-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail 125 9 mm x 165 mm, Item Number 816709165 Product Usage: Intended for the fixation of fractures
  • Z-0132-2019 · Aug 8, 2018 · Terminated
    Affixus Hip Fracture Nail Right 125 11 mm x 340 mm, Item Number 814311340 Product Usage: Intended for the fixation of fractures

…and 234 more.

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page