AFFIXUSHIP FRACTURE NAIL - RIGHTRecalls & reported problems across all models
Biomet Orthopedics, LLC · 39 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 37
We found 39 models of AFFIXUSHIP FRACTURE NAIL - RIGHT by Biomet Orthopedics, LLC in the FDA's device registry (GUDID). Together they are named in 6 adverse event reports in the FDA's MAUDE database, received between Feb 2, 2021 and Jan 17, 2024 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
37 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
AFFIXUSHIP FRACTURE NAIL - RIGHT recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0484-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0483-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0482-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0481-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0480-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0479-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0478-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0477-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0476-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0475-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0474-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0473-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0472-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0471-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0470-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0469-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0468-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0467-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0466-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0465-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0464-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0463-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0462-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0461-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0460-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0459-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0458-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0457-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0456-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Class IITerminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0455-2018 · Root cause: Process change control · Recall details · FDA record ↗
…and 7 older recalls, listed on each model's page.
Most reported AFFIXUSHIP FRACTURE NAIL - RIGHT problems
Device problems
- Fracture5
- Mechanical Jam1
Patient problems
- Failure of Implant2
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Pain1
AFFIXUSHIP FRACTURE NAIL - RIGHT models
39 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-280ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-420ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-460ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-460ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-13-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-13-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-420ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-420ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-13-280ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-13-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-13-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-13-460ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-11-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
Recent AFFIXUSHIP FRACTURE NAIL - RIGHT adverse event reports
The 6 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJan 17, 2024Fracture; Fracture
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL HIP FRACTURE FIXATION PROCEDURE. SUBSEQUENTLY, THREE (3) MONTHS POST-IMPLANTATION, THE PATIENT PRESENTED WITH PAIN AND MOBILITY DIFFICULTIES. DIAGNOSTIC IMAGES REVEALED A FRACTURE OF THE IMPLANTED FEMORAL FIXATION NAIL. THE PATIENT UNDERWENT REVISION SURGERY OF THE AFFECTED NAIL SIX (6) MONTHS POST-IMPLANTATION WITHOUT REPORTED COMPLICATION. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 18528354 ↗MalfunctionJan 3, 2024Fracture; Fracture
IT WAS REPORTED THAT THE NAIL BROKE APPROXIMATELY FOUR MONTHS POST-INITIAL IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 18436484 ↗InjuryJun 20, 2022Fracture; Fracture
IT WAS REPORTED THAT A PATIENT HAD AN INITIAL SURGERY APPROXIMATELY 2 YEARS AGO, SHE HAS SINCE HAD A RIGHT BROKEN AFFIXUS HIP FRACTURE NAIL WHICH HAD TO BE REMOVED AND THE FRACTURE FIXED WITH A SYNTHES TROCHANTERIC HOOK PLATE. THE ORIGINAL FRACTURE WAS A SUB TROCHANTERIC FRACTURE EXTENDING DISTALLY TOWARDS THE MID SHAFT. A CABLE CAN BE SEEN ON THE X-RAY AT THE TIME OF REMOVING THE NAIL SHOWING SOME CALLOUS FORMATION AROUND THE MID TO PROXIMAL SECTION OF BONE AND LOOKS TO HAVE HEALED. THE NAIL HAD BROKEN AT THE CIRCLE SECTION WHERE THE LAG SCREW GOES THROUGH THE NAIL AND INTO THE FEMORAL NECK AND ACETABULA HEAD. THE FEMUR HAS FRACTURED FROM THE LATERAL SIDE APPROXIMATELY 8CM BELOW THE GREATER TROCHANTER AND EXTENDED MEDIALLY TOWARDS THE LESSER TROCHANTER AND INFERIOR FEMORAL NECK. THERE WAS NO DISCERNABLE REASON AS TO WHY THE NAIL HAD BROKEN. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.
Patient problems: Bone Fracture(s); Failure of Implant
MAUDE report 14743667 ↗InjuryApr 8, 2022Fracture; Fracture
NO ADDITIONAL EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 14052340 ↗MalfunctionNov 19, 2021Mechanical Jam; Mechanical Jam
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12844258 ↗InjuryFeb 2, 2021Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL PROCEDURE ON AN UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO FRACTURE OF THE AFFIXUS NAIL. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Code Available
MAUDE report 11265856 ↗
Other Biomet Orthopedics, LLC products
- AFFIXUSHIP FRACTURE NAIL - LEFT27 models · 7 reports · 24 recalls
- Compress Device71 models · 99 reports · 22 recalls
- Arcos Modular Revision Hip System71 models · 191 reports · 19 recalls
- Comprehensive Reverse Shoulder57 models · 895 reports · 3 recalls
- Vanguard Knee System147 models · 306 reports · 13 recalls
- STAGEONE17 models · 0 reports · 18 recalls
- StageOne Select15 models · 2 reports · 15 recalls
- Comprehensive Shoulder System68 models · 191 reports · 11 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page