Skip to content

LEVEL ONE CMF 25-878-05-91Recalls & reported problems

KLS-Martin L.P. · SCREW, MAXDRIVE, DRILL FREE, TI-6AL-4V · Class 2 · Implant

All models of LEVEL ONE CMF (150)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00888118047694), received between Dec 15, 2016 and Aug 27, 2025 — classified as 14 injuries.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Osseointegration Problem4
  • Adverse Event Without Identified Device or Use Problem3
  • Component Falling1
  • Failure to Osseointegrate1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Loss of Osseointegration1

Patient problems

  • Impaired Healing6
  • Device Embedded In Tissue or Plaque3
  • Inadequate Osseointegration3
  • Foreign Body In Patient2
  • Failure of Implant1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    17 DAYS AFTER IMPLANT AND DURING DISTRACTION PHASE, DISTRACTOR SCREWS BECAME LOOSENED FROM 8-MONTH-OLD PATIENT'S BONE. SCREWS WERE UNDAMAGED AND INTACT WITH DISTRACTOR CONSTRUCT ON EXPLANT. THEY WERE REPLACED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22902745 ↗
  • InjuryMay 23, 2025Osseointegration Problem; Osseointegration Problem

    IMAGING REVEALED PATIENT BONE RESORPTION AT MARGINS OF DEVICE TWELVE MONTHS AFTER IMPLANTATION AND CONSEQUENT LOOSENED FIXATION SCREWS. THIS WAS CONFIRMED AT EXPLANT. SCREWS WERE REMOVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22078358 ↗
  • InjuryMay 25, 2022Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    IT WAS REPORTED THAT SCREWS BECAME LOOSE DUE TO PATIENT GRINDING THEIR TEETH. THEY WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 14498335 ↗
  • InjuryMar 30, 2021Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THE SCREWS BECAME LOOSE AND WERE REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 11591066 ↗
  • InjuryJun 19, 2020Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT SCREWS BECAME LOOSE AND WERE REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 10174531 ↗
  • InjuryDec 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED SCREWS SECURING A PLATE WERE REMOVED DUE TO THE PATIENT HAVING TROUBLE WITH EYE MOBILITY UNRELATED TO THE IMPLANTS.

    Patient problems: Impaired Healing

    MAUDE report 9477035 ↗
  • InjuryFeb 5, 2019Break; Break

    IT WAS REPORTED THAT THE SCREWS BROKE, AND A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8306705 ↗
  • InjuryNov 5, 2018Break; Break

    IT WAS REPORTED A SCREW BROKE DURING IMPLANTATION. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8038588 ↗
  • InjuryAug 24, 2018Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT SCREWS BECAME LOOSE. THE LOOSE SCREWS WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 7812527 ↗
  • InjuryMay 16, 2018Break; Break

    IT WAS REPORTED A SCREW HEAD BROKE OFF. THE THREADS OF THE SCREW REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7518481 ↗
  • InjuryMar 6, 2018Failure to Osseointegrate; Failure to Osseointegrate

    IT WAS REPORTED THE PLATES AND SCREWS WERE REMOVED DUE TO LOOSE SCREWS.

    Patient problems: Impaired Healing; Inadequate Osseointegration

    MAUDE report 7317587 ↗
  • InjuryFeb 16, 2017Break

    CHEIF ORAL SURGEON'S RESIDENT BROKE THE HEAD OF THE DRILL FREE SCREW WHILE IMPLANTING IN THE MIDFACE REGION. THERE WERE NO CONSEQUENCES TO THE PATIENT REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 6335643 ↗
  • InjuryJan 31, 2017Break; Component Falling

    THE DOCTOR WAS REPAIRING AN INFRAORBITAL RIM FRACTURE IN THE PATIENT. AFTER DRILLING A PILOT HOLE, SHE THEN PROCEEDED TO PLACE THE SCREW INTO THE PREDRILLED HOLE. THE SCREW SNAPPED OFF SLIGHTLY BEFORE THE SCREW HEAD REACHED THE PLATE. THE BODY OF THE SCREW REMAINS IN THE PATIENT. NO REPORTED INJURIES OCCURRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 6291327 ↗
  • InjuryDec 15, 2016Break

    THE SCREW BROKE OFF INTO THE PATIENTS BONE. A PORTION OF THE SCREW WAS LEFT IN THE BONE DUE TO BEING UNABLE TO REMOVE IT.

    Patient problems: Foreign Body In Patient

    MAUDE report 6177586 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page