LEVEL ONE CMF 25-878-07-91Recalls & reported problems

KLS-Martin L.P. · SCREW, MAXDRIVE, DRILL FREE, TI-6AL-4V · Class 2 · Implant

All models of LEVEL ONE CMF (150)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888118047731), received between Apr 4, 2016 and Dec 19, 2025 — classified as 9 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Fracture2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Loss of Osseointegration1
  • Material Integrity Problem1
  • Osseointegration Problem1

Patient problems

  • No Consequences Or Impact To Patient4
  • Device Embedded In Tissue or Plaque2
  • Impaired Healing2
  • Failure of Implant1
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SCREWS REQUIRED REPOSITIONING DUE TO UNKNOWN REASONS. THEY WERE REMOVED AND NEW SCREWS WERE IMPLANTED.

    Patient problems: Insufficient Information

    MAUDE report 23855154 ↗
  • InjuryMay 28, 2021Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED SCREWS BECAME LOOSE AND WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 11901248 ↗
  • InjurySep 10, 2018Break; Break

    IT WAS REPORTED THAT A SCREW BROKE DURING IMPLANTATION. A PORTION OF THE SCREW REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7860243 ↗
  • InjuryAug 24, 2018Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT SCREWS BECAME LOOSE. THE LOOSE SCREWS WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 7812532 ↗
  • MalfunctionJun 29, 2018Break; Break

    IT WAS REPORTED A SCREW BROKE OFF IN THE PEEK IMPLANT. THE BROKEN PORTION OF THE SCREW LEFT IN THE IMPLANT WAS REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7651585 ↗
  • MalfunctionJun 29, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT'S X-RAY SHOWED DISPLACED PLATES AND SCREWS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7651579 ↗
  • InjuryFeb 13, 2018Fracture; Fracture

    SCREW HEAD BROKE OFF DURING IMPLANTATION AND THE SHAFT OF THE SCREW REMAINS IMPLANTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7267582 ↗
  • InjuryNov 3, 2017Fracture; Fracture

    SCREW HEAD BROKE OFF AT THE THREAD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6999121 ↗
  • InjuryApr 26, 2017Sticking; Material Integrity Problem

    WHEN THE DOCTOR WAS PUTTING THE PLATE ON TO THE SEGMENTALIZED FIBULA BONE AFTER PRE-DRILLING THE HOLES, THE 7 MM SCREWS PARTIALLY THREADED BEFORE BECOMING STUCK IN THE BONE. INCREASED FORCE ON THE SCREWDRIVER TO MOBILIZE SCREWS CAUSED THE SCREW HEADS TO STRIP AND THEY BECAME UNUSABLE. DOCTOR WAS FORCED TO REMOVE THE SCREWS WITH A WIRE-DRIVER, AND ONE OF THE SCREW HEADS BROKE OFF FROM THE THREADS. THE DOCTOR HAD TO MOVE THE PLATE TO A NON-IDEAL LOCATION SLIGHTLY BELOW THE PREVIOUS SCREW HOLES IN ORDER TO FIXATE THE PLATE WHILE AVOIDING THE STRIPPED SCREW BODY. PATIENT IS DOING WELL AND NO OTHER ADVERSE EFFECTS FROM THIS INCIDENT HAVE BEEN REPORTED AT THIS POINT.

    Patient problems: Failure of Implant

    MAUDE report 6522899 ↗
  • InjuryJan 6, 2017Break

    ON (B)(6) 2016 DURING A PROCEDURE TWO MAXDRIVE SCREW HEADS BROKE OFF WHEN PLACED IN THE CRANIUM. REMAINDER OF THE SCREWS WERE LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 6230646 ↗
  • InjuryApr 4, 2016Break

    A SCREW HEAD STRIPPED DURING PLACEMENT. THE SCREW REMAINS IN PATIENT'S BONE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 5544524 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page