Skip to content

LEVEL ONE CMF 25-879-07-91Recalls & reported problems

KLS-Martin L.P. · SCREW, MAXDRIVE, DRILL FREE, TI-6AL-4V · Class 2 · Implant

All models of LEVEL ONE CMF (150)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
19
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00888118047793), received between Dec 6, 2016 and May 20, 2025 — classified as 19 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem9
  • Adverse Event Without Identified Device or Use Problem3
  • Break3
  • Loss of Osseointegration2
  • Fracture1
  • Installation-Related Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Impaired Healing11
  • Inadequate Osseointegration2
  • No Consequences Or Impact To Patient2
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Musculoskeletal problem1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 20, 2025Installation-Related Problem; Installation-Related Problem

    A CMF SCREW FIXATING A PLATE WAS FOUND TO HAVE BEEN INSTALLED ALONG EDGE OF BONE, DAMAGING CORTEX. IT WAS REMOVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22054658 ↗
  • InjuryNov 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CMF SCREWS IMPLANTED IN A RESCUE REVISION PROCEDURE ON PATIENT WITH KNOWN POOR BONE QUALITY BECAME LOOSE AFTER IMPLANT. EXPLANTED SCREWS WERE INTACT AND UNDAMAGED. PATIENT WAS RE-PLATED IN A SECOND REVISION PROCEDURE.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 20698675 ↗
  • InjuryNov 17, 2023Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED A SCREW HAD BECOME LOOSE POSSIBLY DUE TO PATIENT NONCOMPLIANCE. IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 18165254 ↗
  • InjuryApr 5, 2023Osseointegration Problem; Osseointegration Problem

    SCREWS USED TO SECURE A MANDIBULAR IMPLANT TO A PATIENT'S MANDIBLE WERE FOUND TO BE TOO SHORT TO SECURE THE IMPLANT TO THE BONE AND WERE REMOVED AND REPLACED WITH LONGER SCREWS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16686764 ↗
  • InjuryOct 19, 2022Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT A SINGLE SCREW WAS FOUND TO BE LOOSE. ALL OTHER SCREWS AND FIXATION PLATE REMAINED SECURE. DOCTOR ELECTED TO RE-PLATE THE PATIENT.

    Patient problems: Impaired Healing

    MAUDE report 15636012 ↗
  • InjurySep 30, 2022Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT DRILL FREE SCREWS PREVIOUSLY IMPLANTED INTO A PATIENT HAD BECOME LOOSE AND THE PHYSICIAN ELECTED TO REMOVE THEM.

    Patient problems: Insufficient Information

    MAUDE report 15509458 ↗
  • InjuryDec 20, 2021Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT TWO SCREWS BECAME LOOSE DUE TO PATIENT'S ANATOMY. THEY WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 13044594 ↗
  • InjuryDec 2, 2021Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED SCREWS WERE REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 12919850 ↗
  • InjuryOct 28, 2020Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    IT WAS REPORTED THE SCREWS PULLED FROM THE BONE DUE TO PATIENT NON-COMPLIANCE. PRODUCT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 10748555 ↗
  • InjuryOct 15, 2020Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED SCREWS BECAME LOOSE DUE TO PATIENT NONCOMPLIANCE. SCREWS WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 10683436 ↗
  • InjuryMar 26, 2020Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED BY THE PHYSICIAN THAT THE PATIENT WAS NOT COMPLIANT WITH PHYSICIAN'S INSTRUCTIONS.

    Patient problems: Impaired Healing

    MAUDE report 9885077 ↗
  • InjuryFeb 19, 2020Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED LOOSE SCREWS WERE REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 9726395 ↗
  • InjuryOct 14, 2019Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THE SCREWS CAME LOOSE. THEY WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 9183699 ↗
  • InjuryFeb 25, 2019Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT SCREWS CAME LOOSE, AND THE PRODUCT WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 8367376 ↗
  • InjuryOct 29, 2018Break; Break

    IT WAS REPORTED A SCREW BROKE DURING IMPLANTATION. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8016679 ↗
  • InjuryAug 17, 2018Break; Break

    IT WAS REPORTED THAT THE SCREW BROKE DURING IMPLANTATION. THE SCREW WAS REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7793988 ↗
  • InjuryJun 29, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE SCREWS WERE REMOVED AND REPLACED DUE TO REPOSITIONING OF THE PLATE.

    Patient problems: Impaired Healing

    MAUDE report 7651586 ↗
  • InjuryAug 16, 2017Fracture; Fracture

    WHILE THE SCREW WAS BEING IMPLANTED INTO THE PLATE WITH A RIGHT ANGLE SCREWDRIVER, THE SCREW HEAD BROKE OFF. THE BODY OF THE SCREW WAS LEFT IN THE PATIENT'S MANDIBLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6796757 ↗
  • InjuryDec 6, 2016Break

    SURGEON WAS PLATING A LEFT ANGLE FRACTURE. THE SURGEON PRE-DRILLED A PILOT HOLE USING A RIGHT ANGLE DRILL, AND THEN PROCEEDED TO SCREW THE IMPLANT WITH THE DRILL FREE SCREW. THE SCREW HEAD BROKE OFF WHILE BEING IMPLANTED INTO THE PATIENTS MANDIBLE. NO CONSEQUENCES TO THE PATIENT DUE TO THE SCREW BREAKING OFF.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6147420 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates