Skip to content

LEVEL ONE CMF 50-774-20-09Recalls & reported problems

KLS-Martin L.P. · PLATE, RECON, TLTS, STR, 2.0-2.5 MM SCREW, CP TITANIUM · Class 2 · Implant

All models of LEVEL ONE CMF (151)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888118078377), received between Mar 20, 2017 and Oct 4, 2024 — classified as 7 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Noise, Audible1

Patient problems

  • Impaired Healing4
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Nonunion/Delayed-union Bone Fracture1
  • Pain1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 4, 2024Break; Break

    A RECON PLATE INSTALLED AS A TEMPORARY LOAD SHARING DEVICE DURING CONSOLIDATION OF MANDIBLE BONE GRAFT IN 84-YEAR-OLD PATIENT BROKE AFTER THE UNDERLYING BONE GRAFT FAILED.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 20370075 ↗
  • InjuryFeb 14, 2023Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PLATE BROKE. IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 16375243 ↗
  • InjuryDec 14, 2022Break; Break

    IT WAS REPORTED THAT A PLATE BROKE. IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 15978845 ↗
  • InjurySep 23, 2022Break; Break

    IT WAS REPORTED A PLATE BROKE LIKELY DUE TO A PATIENT CONDITION AND WAS REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 15473109 ↗
  • InjuryJan 20, 2021Break; Break

    IT WAS REPORTED A PLATE BROKE AND WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 11202783 ↗
  • InjurySep 30, 2019Break; Break

    IT WAS REPORTED THE PLATE BROKE. IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 9137878 ↗
  • MalfunctionJun 29, 2018Break; Break

    IT WAS REPORTED THAT DURING A POST-OP X-RAY THE SURGEON NOTICED THE PLATE WAS BROKEN. PATIENT'S BONE HEALED WELL, NO PLAN AT THIS TIME TO REMOVE BROKEN PLATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7651570 ↗
  • InjuryMar 20, 2017Fracture; Noise, Audible

    PATIENT HAD A PATHOLOGIC FRACTURE DUE TO MANDIBULAR ATROPHY. PATIENT WAS TREATED WITH RADIATION FOR A CANCEROUS MANDIBULAR LESION IN THE SAME AREA PRIOR TO THE FRACTURE. PATIENT HAD UNDERGONE A SERIES OF HYPERBARIC OXYGEN THERAPY PRIOR TO SURGERY. UPON COMPLETION OF RADIATION THE PATIENT HAD COMPLAINED OF TENDERNESS ON THE LEFT SIDE OF HIS JAW. X-RAY WAS TAKEN WHICH DEMONSTRATED A PATHOLOGIC FRACTURE OF THE LEFT MANDIBLE CLOSE TO THE MANDIBULAR ANGLE. ON (B)(6) 2017 PATIENT UNDERWENT SURGERY AND A TS LOCKING PLATE WAS IMPLANTED. SURGERY WAS SUCCESSFUL WITH REDUCTION AND APPLICATION OF THE PLATE WITH LOCKING SCREWS. SURGERY AND HEALING WAS UNEVENTFUL UNTIL (B)(6) 2017 WHEN THE PATIENT COMPLAINED THAT WHILE HE WAS CHEWING HE HEARD A POPPING NOISE. PATIENT WAS WEARING UPPER AND LOWER DENTURES AT THE TIME THIS OCCURRED. POST-OPERATIVE X-RAY DURING HIS EXAM REVEALED THE LOCKING PLATE HAD FRACTURE AT THE INITIAL FRACTURE SITE. THE DOCTOR PLANNED TO TAKE HIM BACK TO THE OPERATING ROOM AND POSITION OUR EXTERNAL FIXATOR, WHICH WAS TO OCCUR ON (B)(6) 2017. PATIENT CANCELLED SURGERY TO GET A SECOND OPINION. PLATE STILL REMAINS IN PATIENT. DOCTOR REPORTED PATIENT IS NOT IN ANY PAIN AT THIS TIM

    Patient problems: Pain

    MAUDE report 6419026 ↗

FDA clearance and classification

The FDA regulates product code MQN (external mandibular fixator and/or distractor) under 21 CFR 872.4760 as a Class II device. 21 CFR 872.4760 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LEVEL ONE CMF 50-774-20-09

Has the LEVEL ONE CMF 50-774-20-09 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888118078377), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LEVEL ONE CMF 50-774-20-09?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Mar 20, 2017 and Oct 4, 2024. In those reports, the most common specific device problems are “Break” (7), “Fracture” (1) and “Noise, Audible” (1); the most common patient problems are “Impaired Healing” (4), “Nonunion/Delayed-union Bone Fracture” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LEVEL ONE CMF 50-774-20-09?

None of the reports are classified as deaths, 7 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LEVEL ONE CMF 50-774-20-09 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the LEVEL ONE CMF 50-774-20-09, and is it still sold?

The FDA's device database lists KLS-Martin L.P. as the company responsible for it (brand LEVEL ONE CMF, model 50-774-20-09). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices

All EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page