DISTRACTION, INTERNAL 51-402-16-09Recalls & reported problems
KLS-Martin L.P. · TUBE, LEFORT 1, ADJ MIDFACE, BUTTRESS, ADJ LENGTH, TI-6AL-4V · Class 2 · Implant
All models of DISTRACTION, INTERNAL (67)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888118080776), received between Jan 6, 2017 and Feb 16, 2017 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break2
- Detachment Of Device Component1
- Positioning Problem1
- Unintended Movement1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2017Break
ON (B)(6) 2017 PATIENT SNAPPED BUTTRESS TUBE WHILE YAWNING. THE DISTRACTOR WAS REMOVED AND REPLACED WITH PLATES. PATIENT WAS INVOLVED IN PREVIOUSLY REPORTED EVENT UNDER MDR 9610905-2016-00059.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6335669 ↗MalfunctionFeb 1, 2017Detachment Of Device Component; Unintended Movement
DISTRACTOR FOOT PLATE BECAME DETACHED FROM THE DISTRACTOR TUBE AFTER ADVANCEMENT WAS COMPLETED. THE FOOT PLATE REMAINED FIXATED TO THE PATIENT. DOCTOR DID NOT THINK IT WAS NECESSARY TO REMOVE THE DEVICE EARLIER THAN PLANNED. PATIENT HAD REPORTED THAT THEY YAWNED/SNEEZED AND FELT A SHIFT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6293004 ↗InjuryJan 6, 2017Break; Positioning Problem
ON (B)(6) 2016 THE PATIENT'S MOTHER HAD DIFFICULTY MAKING THE DISTRACTOR TURN ON THE RIGHT SIDE. THE PATIENT CAME INTO THE CLINIC ON (B)(6) 2016 WHERE DR. (B)(6) ALSO HAD DIFFICULTY TURNING THE DEVICE ON THE RIGHT SIDE. AS HE WAS ASSERTING FORCE TO THE ACTIVATION ARM VIA THE PATIENT WRENCH, HE HEARD A SNAP. CT CONCLUDED THAT THE BUTTRESS TUBE HAD SNAPPED SUPERIOR TO THE SCREW CONNECTION BETWEEN DISTRACTOR BODY AND BUTTRESS TUBE. PATIENT UNDERWENT SURGERY THE SAME DAY TO REPLACE THE DISTRACTOR BODY AND BROKEN BUTTRESS TUBE. DURING THE PROCEDURE THE DOCTOR RETRACTED THE DISTRACTOR ON THE RIGHT SIDE, AND AGAIN THE BUTTRESS TUBE BROKE IN THE SAME PLACE. BUTTRESS TUBE WAS SWAPPED OUT FOR ANOTHER. THE RESIDENT AND FELLOW THOUGHT THAT WHILE RETRACTING THE DISTRACTORS, THE SIDES BECAME IN EVEN, PUTTING TOO MUCH LOAD ON THE RIGHT SIDE MAKING IT SNAP.
Patient problems: No Code Available
MAUDE report 6230629 ↗
FDA clearance and classification
The FDA regulates product code MQN (external mandibular fixator and/or distractor) under 21 CFR 872.4760 as a Class II device. 21 CFR 872.4760 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DISTRACTION, INTERNAL 51-402-16-09
Has the DISTRACTION, INTERNAL 51-402-16-09 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888118080776), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DISTRACTION, INTERNAL 51-402-16-09?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jan 6, 2017 and Feb 16, 2017. In those reports, the most common specific device problems are “Break” (2), “Detachment Of Device Component” (1) and “Positioning Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DISTRACTION, INTERNAL 51-402-16-09?
None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DISTRACTION, INTERNAL 51-402-16-09 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the DISTRACTION, INTERNAL 51-402-16-09, and is it still sold?
The FDA's device database lists KLS-Martin L.P. as the company responsible for it (brand DISTRACTION, INTERNAL, model 51-402-16-09). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.