DISTRACTION, INTERNAL 51-540-25-09Recalls & reported problems
KLS-Martin L.P. · DIST, LEFORT 1, TSMD, ANTWERP TRANS SINUSOIDAL, RIGID ARM, 1.5-1.8 MM SCREW, TI-6AL-4V · Class 2 · Implant
All models of DISTRACTION, INTERNAL (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888118081933), received between Oct 14, 2016 and Oct 14, 2016 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 14, 2016Device Dislodged or Dislocated
DEVICE ACTIVATION ARM SNAPPED OFF DURING THE DISTRACTION PROCESS. REPLACED BOTTOM HALF OF THE DISTRACTOR WITH ACTIVATION ARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6031542 ↗
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page