Skip to content

LEVEL ONE THORACIC, STERILE 24-015-25-71Recalls & reported problems

KLS-Martin L.P. · PLATE, RIB, LOCKING, X SHP, 2.3 MM SCREW, CP TITANIUM · Class 2 · Implant

All models of LEVEL ONE THORACIC, STERILE (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00888118092540), received between Sep 2, 2016 and Mar 1, 2024 — classified as 16 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break17
  • Fracture1
  • Use of Device Problem1

Patient problems

  • Impaired Healing12
  • No Known Impact Or Consequence To Patient3
  • Failure of Implant1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 1, 2024Break; Break

    IT WAS REPORTED A RIB PLATE BROKE. IT WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18815265 ↗
  • InjuryDec 14, 2022Break; Break

    IT WAS REPORTED A THORACIC RIB PLATE BROKE. IT WAS REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 15978853 ↗
  • InjuryOct 14, 2020Break; Break

    IT WAS REPORTED PLATES BROKE AND WERE REMOVED.

    Patient problems: Impaired Healing; Pain

    MAUDE report 10676567 ↗
  • InjuryJul 9, 2020Break; Break

    IT WAS REPORTED A PLATE BROKE AND WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 10255087 ↗
  • InjuryJan 2, 2020Break; Break

    IT WAS REPORTED THE PLATES BROKE. THEY WERE REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 9537654 ↗
  • MalfunctionNov 15, 2019Break; Break

    IT WAS REPORTED PLATES BROKE AND REMAINS IMPLANTED. THE PATIENT DECIDED NOT TO HAVE THE PLATES REMOVED AT THIS TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9330096 ↗
  • InjuryMar 18, 2019Break; Break

    IT WAS REPORTED THAT PLATES BROKE. EXPECTED TO BE REMOVED ON AN UNKNOWN DATE.

    Patient problems: Impaired Healing

    MAUDE report 8428562 ↗
  • MalfunctionSep 14, 2018Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT A PLATE BROKE. THERE ARE NO PLANS TO REMOVE THE PLATE AT THIS TIME. IFU (B)(4) STATES "THE SURGEON MUST ENSURE THE FRACTURED SEGMENTS HAVE BONE TO BONE CONTACT. ANY BRIDGING OF A CONTINUITY DEFECT MUST BE COMBINED WITH A BONE GRAFT".

    Patient problems: Impaired Healing

    MAUDE report 7877975 ↗
  • InjurySep 10, 2018Break; Break

    IT WAS REPORTED THAT A PLATE BROKE AND REMAINS IMPLANTED.

    Patient problems: Impaired Healing

    MAUDE report 7860232 ↗
  • InjuryAug 13, 2018Break; Break

    IT WAS REPORTED THAT A PLATE BROKE AND REMAINS IMPLANTED.

    Patient problems: Impaired Healing

    MAUDE report 7777915 ↗
  • InjuryAug 8, 2018Break; Break

    IT WAS REPORTED THAT A PLATE BROKE AND REMAINS IMPLANTED.

    Patient problems: Impaired Healing

    MAUDE report 7764734 ↗
  • InjuryJun 29, 2018Break; Break

    IT WAS REPORTED THE PLATE BROKE WHILE PATIENT WAS COUGHING.

    Patient problems: Impaired Healing

    MAUDE report 7651598 ↗
  • InjuryJun 8, 2018Break; Break

    IT WAS REPORTED THAT A PLATE BROKE AND WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 7584213 ↗
  • InjuryFeb 15, 2018Break; Break

    IT WAS REPORTED THAT THE PATIENT WAS TREATED FOR RIB FRACTURES BUT WAS NON-COMPLIANT WITH POST-OP INSTRUCTIONS. APPROXIMATELY FOUR WEEKS POST-OP, THE PATIENT CHANGED A CAR TIRE. AFTER THIS EVENT, THE PATIENT FELT MOVEMENT AND SOMETHING WRONG IN THE LOCATION OF THE FRACTURES. UPON EXAMINATION, IT WAS DETERMINED THAT BOTH PLATES HAD FRACTURED. IMPLANTS WERE REMOVED, AND RIBS WERE REPAIRED WITH IMPLANTS FROM ANOTHER COMPANY.

    Patient problems: Impaired Healing

    MAUDE report 7272837 ↗
  • InjuryFeb 15, 2018Break; Break

    IT WAS REPORTED THAT A RIB PLATE BROKE. PRODUCT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 7272829 ↗
  • InjuryMar 31, 2017Fracture

    RIB PLATE FRACTURED SOMETIME AFTER IMPLANTATION AND WAS REMOVED ON (B)(6) 2017.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6449178 ↗
  • InjuryFeb 16, 2017Break

    ONE MONTH POST-OPERATION PATIENT HAD A SCAN AND ALL IMPLANTED PLATES LOOKED FINE. PATIENT CAME BACK TO THE HOSPITAL IN (B)(6) 2016 FOR ANOTHER PROCEDURE AND IT WAS DISCOVERED THAT ONE OF THE PLATES WERE BROKEN. WHILE SHE WAS INPATIENT FOR THE OTHER PROCEDURE THE OTHER TWO PLATES WERE REPORTED TO BREAK. THE BROKEN PLATES WERE REMOVED AND REPLACED ON (B)(6) 2017. IT WAS REPORTED THE PATIENT WAS A SMOKER UP UNTIL THE INITIAL SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 6335596 ↗
  • InjurySep 2, 2016Break

    A PLATE BROKE AFTER IMPLANTATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5924979 ↗