DISTRACTION, INTERNAL 50-347-05-61Recalls & reported problems
KLS-Martin L.P. · SCREW, MAXDRIVE/HEX, ZURICH, DRILL FREE, 2.0 MM HEX DRIVER, TI-6AL-4V · Class 2 · Implant
All models of DISTRACTION, INTERNAL (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888118113399), received between Aug 10, 2023 and Aug 10, 2023 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AFTER IMPLANTATION, 3.5 MM FIXATION SCREWS WERE FOUND TO BE TOO SHORT FOR APPLICATION. THEY WERE REMOVED AND REPLACED WITH 5.0 MM SCREWS.
Patient problems: Insufficient Information
MAUDE report 17497660 ↗
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page