Skip to content

DISTRACTION, INTERNAL 50-347-07-61Recalls & reported problems

KLS-Martin L.P. · SCREW, MAXDRIVE/HEX, ZURICH, DRILL FREE, 2.0 MM HEX DRIVER, TI-6AL-4V · Class 2 · Implant

All models of DISTRACTION, INTERNAL (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888118113405), received between Aug 23, 2023 and May 2, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration2
  • Adverse Event Without Identified Device or Use Problem1
  • Osseointegration Problem1

Patient problems

  • Inadequate Osseointegration3
  • Neonatal Deformities1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 2, 2025Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED DISTRACTOR SCREWS BECAME LOOSE FROM COMPROMISED BONE DURING DISTRACTION. THEY WERE REMOVED AND REPLACED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21950210 ↗
  • InjuryOct 31, 2024Loss of Osseointegration; Loss of Osseointegration

    PHYSICIAN REPORTED A DISTRACTOR SCREW BECAME LOOSE FROM PATIENT'S BONE DUE TO POOR BONE QUALITY. PHYSICIAN SPECIFICALLY NOTED DEVICE DID NOT FAIL OR HAVE ANY ISSUES. DEVICE WAS REMOVED AND REPLACED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20577387 ↗
  • InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SEVERED COMPROMISED BONE FRAGMENT INTO WHICH DISTRACTOR FOOTPLATE SCREWS WERE IMPLANTED DID NOT SUPPORT DISTRACTION.

    Patient problems: Neonatal Deformities

    MAUDE report 18584282 ↗
  • InjuryAug 23, 2023Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT SCREWS SECURING A DISTRACTOR BECAME LOOSE FROM BONE. THE PHYSICIAN REMOVED THE SCREWS AND REPLACED THEM WITH THE SAME SIZE AND MODEL SCREWS WHILE LEAVING THE DISTRACTOR IN PLACE. THE SCREWS WERE NOTED TO BE INTACT AND UNDAMAGED UPON EXPLANT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17607193 ↗

Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices

All EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page