HeartWare® Battery 1650DE (UDI 00888707000369)Recalls & reported problems
Heartware Inc. · Class 3 · Implant
All models of HeartWare Battery (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This HeartWare battery powers the HVAD heart pump, which is implanted to help the heart pump blood. Of its 1,126 reports, 5 describe injuries and 1,121 describe malfunctions, with no deaths. Most reports describe batteries that drain quickly or fail to charge, and power disconnect alarms. The most recent report is from January 2024. 1 open recall, started in May 2018, concerns brief connection interruptions that cause unexpected switches to backup power and beeping.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,126
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,126 adverse event reports that list this exact device (DI 00888707000369), received between Jan 12, 2016 and Jan 16, 2024 — classified as 5 injuries, 1,121 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (1,040 reports).
1 FDA recall record name this device's identifier, 1 still open.
59 other recalls cover devices cleared under the same FDA submission (P100047); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 2, 2018Open, Classified
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.
Z-1903-2018 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Battery Problem1,040
- Connection Problem52
- Power Problem46
- Material Integrity Problem32
- Display or Visual Feedback Problem12
- Material Deformation9
- Charging Problem8
- Failure to Charge7
Patient problems
- No Known Impact Or Consequence To Patient831
- No Consequences Or Impact To Patient158
- No Clinical Signs, Symptoms or Conditions143
- No Patient Involvement19
- Complaint, Ill-Defined1
- Dizziness1
- Loss of consciousness1
- Missing Value Reason1
Recent adverse event reports
The 25 most recent of 1,126 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2024Battery Problem; Battery Problem
A BATTERY WAS RETURNED TO THE MANUFACTURER AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT C OMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18519143 ↗MalfunctionOct 17, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY HAS OVER 500 CYCLES AND IT IS DEPLETING ON A VERY SHORT TIME. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17950845 ↗MalfunctionAug 24, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT A POWER DISCONNECT ALARM OCCURRED WHILE THE BATTERY WAS CONNECTED PROPERLY TO THE CONTROLLER. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17612593 ↗MalfunctionAug 22, 2023Battery Problem; Battery Problem
IT WAS FURTHER REPORTED THAT THE BATTERY WAS REMOVED FROM SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17599351 ↗MalfunctionAug 17, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERIES WERE NOT CHARGING. BOTH BATTERIES HAVE BEEN REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17565449 ↗MalfunctionAug 15, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE BATTERY HAD A BROKEN AND CRACKED CABLE. THE BATTERY WAS REMOVED FROM USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17549380 ↗MalfunctionAug 10, 2023Battery Problem; Battery Problem
I WAS REPORTED THAT THE BATTERY DID NOT PROVIDE POWER, EVEN THOUGH IT WAS CONNECTED PROPERLY TO THE CONTROLLER A POWER DISCONNECT ALARM OCCURRED. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17498505 ↗MalfunctionAug 1, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THREE BATTERIES DEPLETED VERY FAST. IT WAS NOTED THAT THE BATTERIES HAD OVER 500 CHARGE CYCLE COUNTS. THE BATTERIES WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17434954 ↗MalfunctionJul 12, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE BATTERY EXHIBITED UNSPECIFIED DAMAGE. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17305820 ↗MalfunctionJun 30, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY DOES NOT PROVIDE POWER. ALTHOUGH THOUGH THE BATTERY WAS CONNECTED TO THE CONTROLLER PROPERLY A POWER DISCONNECT ALARM OCCURRED. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17238357 ↗MalfunctionJun 13, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY WAS NOT PROVIDING POWER TO THE CONTROLLER AND THERE WAS A POWER DISCONNECT ALARM EVEN THOUGH THE BA TTERY WAS CONNECTED TO THE CONTROLLER PROPERLY. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17116909 ↗MalfunctionJun 9, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY WAS PROPERLY CONNECT TO THE CONTROLLER BUT DID NOT PROVIDE POWER RESULTING IN A POWER DISCONNECT ALARM OCCURRED. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17098275 ↗MalfunctionJun 1, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY EXHIBITED A COMMUNICATION ERROR. THE BATTERY REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17038435 ↗MalfunctionMay 23, 2023Battery Problem; Battery Problem
IT WAS REPORTED THE BATTERIES DID NOT PROVIDE POWER EVENT THOUGH THEY WERE PROPERLY CONNECTED TO THE CONTROLLER. IT WAS ALSO REPORTED THAT POWER DISCONNECT ALARMS OCCURRED. THE BATTERIES WERE DISCARDED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16986113 ↗MalfunctionMay 23, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT TWO BATTERIES EXHIBITED UNSPECIFIED DAMAGE. THE BATTERIES WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16985555 ↗MalfunctionApr 25, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT FOUR (4) OF THE PATIENT BATTERIES WERE NOT SUPPORTING >3-4 HOURS OF CHARGE. THE BATTERIES WERE REMOVED FROM SER VICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807302 ↗MalfunctionApr 20, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY DID NOT PROVIDE POWER AND EXHIBITED A POWER DISCONNECT ALARM. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16776890 ↗MalfunctionApr 3, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERIES DID NOT PROVIDE POWER AND EXHIBITED POWER DISCONNECT ALARMS WHILE CORRECTLY BEING ATTACHED TO THE CONTROLLER. THE BATTERIES WERE REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16667268 ↗MalfunctionMar 9, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERIES DID NOT PROVIDE POWER ALTHOUGH THEY WERE CONNECTED PROPERLY TO THE CONTROLLER, POWER DISCONNECT ALARMS OCCURRED. THE BATTERIES HAVE BEEN EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16514140 ↗MalfunctionMar 8, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY DOES NOT PROVIDE POWER. IT WAS ALSO REPORTED THAT THE BATTERY WAS PROPERLY CONNECTED TO THE CONTROLLER AND A POWER DISCONNECT ALARM OCCURRED. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16504182 ↗MalfunctionMar 2, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY EXHIBITED A POWER DISCONNECT ALARM AND DID NOT PROVIDE POWER. IT WAS NOTED THAT THE BATTERY WAS CONNECTED TO THE CONTROLLER PROPERLY. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16468578 ↗MalfunctionFeb 28, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THREE BATTERIES WERE EXHIBITING MULTIPLE ISSUES. THE THREE BATTERIES DID NOT PROVIDE POWER AND EXHIBITED MULTIPLE POWER DISCONNECT ALARMS, EVEN WHEN THE BATTERIES WERE PROPERLY CONNECTED TO THE CONTROLLER. THE BATTERIES WERE REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16452826 ↗MalfunctionFeb 15, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERIES EXHIBITED POWER DISCONNECT ALARMS AND WERE NOT PROVIDING POWER WHEN PLUGGED INTO THE CONTROLLER. THE BATTERIES WERE REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16381087 ↗MalfunctionFeb 13, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY DOES NOT PROVIDE POWER TO THE CONTROLLER. EVEN THOUGH THE BATTERY WAS CONNECTED TO THE CONTROLLER PROPERLY, A POWER DISCONNECT ALARM OCCURRED. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16363553 ↗MalfunctionFeb 10, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE BATTERY EXHIBITED POWER CONSUMPTION ISSUE AS IT LAST FOR LESS THAN TWO HOURS. THE BATTERY WAS REMOVED FROM SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16353670 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100047. One submission often covers many models, so check the product and lot details in each record.
- Z-0475-2024 · Oct 31, 2023 · Open, ClassifiedHVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100
- Z-1993-2023 · May 16, 2023 · Open, ClassifiedHeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore Tech, c) Monitor 1521IL Explore Tech, d) Monitor 1521DE Explore Tech
- Z-1512-2022 · Jun 28, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Batteries, Product Number 1650DE
- Z-1266-2022 · May 5, 2022 · Open, ClassifiedHeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
- Z-1207-2022 · Apr 14, 2022 · Open, ClassifiedHeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU
- Z-1105-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420-Controller 2.0
- Z-1107-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, individual pamphlets: HVAD System Emergency Responder Guide (FI-Finnish) HVAD System Emergency Responder Guide (FI-Finnish) H
- Z-1104-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420
- Z-1110-2022 · Mar 30, 2022 · Open, ClassifiedInstructions for Use, Patient Manuals, and Emergency Responder Guides for HeartWare Ventricular Assist Device (HVAD) System. HVAD System. IFUs are distributed as part of the Pump
- Z-1102-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, Pump Implant Kit, Model Number1104
- Z-1103-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1407DE Kit
- Z-1106-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, Model Number 1521DE, MONITOR 1521DE EXPLORE TECH
- Z-0726-2022 · Jan 27, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System
- Z-2478-2021 · Aug 6, 2021 · Open, ClassifiedMedtronic HVAD Pump Implant Kits
- Z-2197-2021 · Jun 3, 2021 · Open, ClassifiedHeartWare HVAD Pump Accessories, REF MCS1753AK
…and 44 more.
Other Ventricular (Assisst) Bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page