Heartware Medtronic HVAD System for cardiac use.… recallZ-1903-2018 · Heartware · initiated May 2, 2018
Class I Open, Classified All recalls posted in May 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heartware or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.
FDA root cause category: Device Design
What the company is doing
The firm notified physicians and healthcare professionals about the device correction by letter on 05/02/2018. The letter explained problem and provided recommendations for effective power source management of the system. Medtronic began notifying VAD Clinicians in the US on May 31, 2018, mainly via 2-day courier, and OUS via locally approved methods, of a newly available field service. A field service for application of a lubricant to the HVAD System power source connectors to mitigate the potential for transient interruptions of the electrical connection between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the HVAD Controller is now available.
Recall details
- Recall number
- Z-1903-2018
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- May 2, 2018
- Posted by the FDA
- May 21, 2018
- Recalling firm
- Heartware
- Quantity
- 175878 devices
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, and Vietnam.
Product description
Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
ALL SERIAL NUMBERS UPN/GTIN: (a) 1400 and 1401: distributed prior to GTN requirement 1403: 00888707000116, 00888707000475 1407: '00888707000291, '00888707000727, '00888707000710, 00888707000734, '00888707000741, '00888707000765, '00888707000758, '00888707000789, '00888707000772, '00888707000499, '00888707000796, '00888707000802, '00888707000819, '00888707000482, '00888707004534, '0888707001670, '00888707001656, '00888707001663, '00888707001670, '00888707001687, '00888707001694, '00888707001700, '00888707001717, '00888707001724, '00888707001731, '00888707001748, '00888707001755, '00888707001762, '00888707002813, '00888707002820, '00888707002837, '00888707002844, '00888707002851, '00888707001472 1420: '00888707000420, '00888707000437, '00888707002660 (b) 1435: '00888707000109 1440: '00888707000260, '00888707001885, '00888707001496, '00888707002745 (c) 1425: '00888707000093 1430: '00888707000307, '00888707000826, '00888707000833, '00888707000840, '00888707000857, '00888707000871, '00888707000864, '00888707000888, '00888707000901, '00888707000505, '00888707000895, '00888707000918, '00888707001489, '00888707001779, '00888707001786, '00888707001793, '00888707001809, '00888707001816, '00888707001823, '00888707001830, '00888707001847, '00888707001854, '00888707001861, '00888707001878, '00888707002738, '00888707002769, '00888707002776, '00888707002783, '00888707002790, '00888707002806, '00888707004817 (d) 1650DE: 00888707000369, 00888707001373, '00888707001588, '00888707001366, '00888707002684 '00888707002646, '00888707002653, '00888707000376
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass1 recall5 deaths898 reports
- HeartWare® Battery 1650Heartware Inc. · Ventricular (Assisst) Bypass3 recalls740 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall130 reports
- HeartWare® Controller AC Adapter 1430USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 reports
- HeartWare® Controller AC Adapter 1425USHeartware Inc. · Ventricular (Assisst) Bypass1 recall36 reports
- HeartWare® Controller AC Adapter 1430USHeartware Inc. · Ventricular (Assisst) Bypass1 recall25 reports
- HeartWare® Controller DC Adapter 1435Heartware Inc. · Ventricular (Assisst) Bypass1 recall21 reports
- HeartWare® Controller DC Adapter 1440Heartware Inc. · Ventricular (Assisst) Bypass1 recall19 reports
- HeartWare® Controller DC Adapter 1440Heartware Inc. · Ventricular (Assisst) Bypass1 recall8 reports
Product lines
- HeartWare BatteryHeartware Inc. · 11 models · 4,241 reports · 5 recalls
- HeartWare Controller AC AdapterHeartware Inc. · 5 models · 97 reports · 3 recalls
- HeartWare Controller DC AdapterHeartware Inc. · 6 models · 59 reports · 1 recall
- HeartWare ControllerHeartware Inc. · 2 models · 2,393 reports · 3 recalls
- HVAD ControllerHeartware Inc. · 3 models · 2,990 reports · 5 recalls
Other devices under the same FDA submission
25 devices cleared under P100047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass4 recalls1 report
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall150 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall109 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page