Arthrex® AR-1223SRecalls & reported problems
Arthrex, Inc. · CORING REAMER, 9MM WITH COLLARED PIN · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867005532), received between Apr 25, 2022 and Nov 3, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Failure to Cut1
- Material Twisted/Bent1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2025Failure to Cut; Failure to Cut
ON 10/22/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA (B)(4) THAT AN AR-1223S 9MM CANNULATED CORING REAMER TREPHINE LOST ITS CUTTING ABILITY AND BECAME STRIPPED. IT WAS NECESSARY TO OPEN ANOTHER TREPHINE TO COMPLETE THE ACL RECONSTRUCTION. NO PATIENT WAS AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452756 ↗MalfunctionOct 15, 2025Positioning Failure; Positioning Failure
ON (B)(6) 2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA (B)(4) THAT AN AR-1223S 9MM CANNULATED CORING REAMER TREPHINE OPENED TOO WIDE AND LOST ITS INTENDED FUNCTION. THIS OCCURRED DURING AN ACL RECONSTRUCTION, AND NO PATIENT WAS AFFECTED. THE PROCEDURE WAS COMPLETED BY OPENING A SECOND TREPHINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23300174 ↗InjuryMay 10, 2023Break; Material Twisted/Bent; Break; Material Twisted/Bent
ON 04/17/2023, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA SEMS THAT AN AR-1223S CORING REAMER BECAME DEFORMED AND BROKE. THIS WAS DISCOVERED DURING A CASE AND BROKE DURING USE. PER FACILITY: "THE PRODUCT WAS USED FOR ARTHROSCOPIC BONE GRAFT HARVESTING FROM THE TALUS ON (B)(6) 2023. DURING USE, THE TIP OF THE REAMER TOUCHED THE SHEATH OF THE CAMERA SCOPE, CAUSING DEFORMATION AND BROKEN WITHIN THE JOINT. AT THIS TIME, IT WAS ALSO CONFIRMED THAT METAL POWDER WAS GENERATED. THE DOCTOR IN CHARGE WAS ABLE TO EXTRACT A BONE GRAFT USING THE PRODUCT AS IT WAS, BUT THE BONE HOLE WAS ENLARGED DUE TO THE DEFORMATION OF THE PRODUCT. NO FURTHER INFORMATION IS AVAILABLE."
Patient problems: Insufficient Information
MAUDE report 16906579 ↗MalfunctionApr 25, 2022Break
ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-1223S CORING REAMER WAS DAMAGED, NO ADDITIONAL INFORMATION WAS PROVIDED. THIS WAS DISCOVERED DURING USE WITH NO IMPACT TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON (B)(6) 2022, THE SALES REPRESENTATIVE STATED THE FOLLOWING INFORMATION VIA PHONE: THIS EVENT OCCURRED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2022. WHEN DRILLING THE TIBIA, THE REAMER BROKE INSIDE THE TIBIA. ALL FRAGMENTS WERE RETRIEVED AND THE CASE WAS COMPLETED SUCCESSFULLY WITH NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14194562 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page