Arthrex® AR-1224SRecalls & reported problems
Arthrex, Inc. · CORING REAMER, 10MM WITH COLLARED PIN · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867005648), received between May 23, 2019 and Mar 1, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation3
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Material Twisted/Bent1
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Bone Fracture(s)2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2024Material Deformation; Material Deformation
ON (B)(6) 2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA SEMS (B)(4) THAT THE SURGEON HAS USED THE CORING REAMER, AR-1224S, IN MANY CASES AND IS FAMILIAR WITH THE TECHNIQUE. HE FOLLOWED THE PROCEDURE TO DRILL THE TIBIAL TUNNEL AND UPON EXITING INTO JOINT, THE TIP OF THE CORING REAMER EXPANDED AND FLOWERED OPEN. THERE ARE MULTIPLE SPLITS IN THE REAMER NEAR THE TIP. HE WAS ABLE TO REMOVE THE DRILL AND NO REMAINS WERE INSIDE THE JOINT. THE ISSUE WAS NOTICED AFTER THE DEVICE EXITED THE CAVITY. THERE WAS NO ADVERSE EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18816265 ↗MalfunctionNov 22, 2023Material Deformation; Material Deformation
ON 10/27/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-1224S CANNULATED CORING REAMER WITH COLLARED PIN, 10 MM, HAD AN ISSUE. WHEN REAMING THE DEVICE, THE TIP FANNED OUT INTO A FLOWER-LIKE FORM INSTEAD OF THE EXPECTED CIRCULAR FINISH. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18192043 ↗MalfunctionNov 20, 2023Material Deformation; Material Deformation
ON (B)(6). 2023 IT WAS REPORTED, BY A SALES REPRESENTATIVE VIA (B)(4). THAT (QTY. 2) OF AN AR-1224S CANNULATED CORING REAMER WITH COLLARED PIN, 10 MM, HAD AN ISSUE. WHEN REAMING THE DEVICE, THE TIP FANNED OUT INTO A FLOWER-LIKE FORM INSTEAD OF THE EXPECTED CIRCULAR FINISH. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18174217 ↗InjuryMay 24, 2021Material Twisted/Bent
IT WAS REPORTED THAT THE METAL AT THE LEADING EDGE OF THE REAMER MUSHROOMED WHEN IT WAS DRILLED INTO THE BONE DURING AN ACL CASE. IT WENT FROM 10 MM TO 24 MM. IT STOPPED BEFORE IT HIT THE JOINT; THERE WAS A HAIRLINE FRACTURE IN THE TIBIA. THE PATIENT WAS MALE, (B)(6) YEARS.
Patient problems: Bone Fracture(s)
MAUDE report 11871773 ↗InjurySep 2, 2020Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED DURING AN ACL RECONSTRUCTION THE TIP OF THE REAMER EXPANDED AND PRODUCED SHARDS OF METAL. A SHAVER WAS USED TO DISTRACT AS MUCH OF THE METAL DEBRIS AS POSSIBLE. THE DOCTOR DID NOT DO AN X-RAY TO INSURE ALL METAL WAS RECOVERED AND THE REP IS NOT 100% CONFIDENT ALL WAS RETRIEVED. THE REST OF THE CASE WAS COMPLETED WITHOUT ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10483716 ↗InjuryMay 23, 2019Break; Smoking
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, WHILE REAMING WITH THE 10MM CORING REAMER, AR-1224S, THE CORING REAMER BEGAN TO SMOKE AND THEN FELL APART. THIS CAUSED A FRACTURE IN THE FEMORAL CONDOYLE. NO FURTHER DETAILS. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION OBTAINED 4/26/2019: FACILITY PROVIDED COPY OF HANDWRITTEN MEDWATCH REPORT (NO REPORT NUMBER LISTED). FACILITY MEDWATCH REPORT STATED THE FOLLOWING: A CORE REAMER SHATTERED RESULTING IN LIFTING A CENTRAL CORE OF THE TIBIAL PLATEAU, CAUSING AN INTRA-ARTICULAR TIBIAL PLATEAU FRACTURE OF THE CENTRAL ASPECT OF THE PLATEAU. THIS REQUIRED SURGICAL REPAIR. ADDITIONAL INFORMATION PROVIDED 5/29/2019: THE SURGEON USED THE CORING REAMER TO BREAK THROUGH THE CORTEX INSTEAD OF AN OVER-SIZED REAMER.
Patient problems: Bone Fracture(s)
MAUDE report 8636742 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page