Skip to content

Arthrex® AR-13200M-055RRecalls & reported problems

Arthrex, Inc. · PLT,LO-PRO,META OPEN WDG,TI,R,5.5MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867009196), received between Nov 9, 2016 and Nov 9, 2016 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Nonunion/Delayed-union Bone Fracture1
  • Pain1
  • Swelling1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 9, 2016Break

    IT WAS REPORTED ON 9/15/2016 BY A PATIENT VIA THE ARTHREX WEBSITE THAT THE PATIENT HAD UNDERGONE FOOT SURGERY (B)(6) 2015. SURGERY WAS LEFT PROXIMAL BUNIONECTOMY IN CONJUNCTION WITH 2ND MTP RECONSTRUCTION FOLLOWING DISLOCATION AND PRIOR PODIATRIC MANAGEMENT. DURING THE SURGERY THE FOLLOWING PARTS WERE IMPLANTED: AR13200M-055R ((B)(4)), AR-13226T (QTY 2) ((B)(4)), AR-8724-16 ((B)(4)), AR-8724-12 ((B)(4)), AR-8724-14 ((B)(4)), AR-8724-18 ((B)(4)), AR-8930-12 ((B)(4)) AND AR-8724-12PT ((B)(4)) ON (B)(6) 2016 PATIENT FOLLOWED UP WITH SURGEON DUE TO RECENT ONSET OF LEFT MID-FOOT PAIN AND SWELLING. MEDICAL RECORDS NOTED THAT UPON EXAMINATION OF THE LEFT FOOT THERE WAS PAIN AT THE 1ST TMT AND PROXIMAL OSTEOTOMY SITE WITH VISIBLE SWELLING. THERE WAS NO ERYTHEMA, DRAINAGE OR OTHER EVIDENCE OF INFECTION. X-RAY RESULTS NOTED THERE IS BROKEN HARDWARE (PROXIMAL BUNION PLATE - AR-13200M-055R) WITH QUESTION OF NON-UNION. SURGEON HAS RECOMMENDED A PROXIMAL LEFT BUNION HARDWARE REMOVAL WITH A PLAN TO PROCEED WITH A LAPIDUS FUSION IF IN FACT THE OSTEOTOMY IS INCOMPLETE. PATIENT REVISION IS TENTATIVELY SCHEDULED FOR EARLY 2017 DUE TO PATIENT LACK OF AVAILABLE TIME OFF FROM WORK.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Swelling

    MAUDE report 6088753 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page