Skip to content

Arthrex® AR-1360C-CPRecalls & reported problems

Arthrex, Inc. · IMPLANT SYSTEM, MPFL, BIOCOMPOSITE · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867014015), received between Oct 31, 2017 and Feb 4, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Appropriate Device Problem Term/Code Not Available1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Difficult to Insert1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Bacterial Infection1
  • Failure of Implant1
  • Foreign Body In Patient1
  • No Code Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Break; Break

    ON 01/13/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT THE ANCHOR FROM AN AR-1360C-CP MPFL BIOCOMPOSITE IMPLANT SYSTEM, BROKE. THE SURGEON DID NOT HAVE THE APPROPRIATE INSTRUMENTATION FOR THE CORRECT ANCHOR PLACEMENT. HALF OF THE SCREW REMAINED IN THE PATIENT; THE OTHER HALF WAS REMOVED. THE CASE WAS COMPLETED BY OPENING A NEW AR-1360C-CP MPFL BIOCOMPOSITE IMPLANT SYSTEM. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION RECEIVED ON 1/22/2025: THIS WAS DISCOVERED DURING A MEDIAL PATELLOFEMORAL LIGAMENT RECONSTRUCTION PROCEDURE. THE 6 MM BIOCOMPOSITE INTERFERENCE SCREW WAS WHAT BROKE IN THE PATIENT. THE AR-1360C-CP MPFL BIOCOMPOSITE IMPLANT SYSTEM WAS NOT MISSING ANY INSTRUMENTATION, HOWEVER, THE SURGEON DID NOT HAVE THE APPROPRIATE DRIVER TO IMPLANT THE INTERFERENCE SCREW, WHICH ENDED UP BREAKING IN THE PATIENT. THE CASE WAS ACTUALLY COMPLETED BY OPENING AN AR-2324BCC BIOCOMPOSITE SWIVELOCK C IN PLACE OF THE INTERFERENCE SCREW. THE CASE WAS DELAYED BETWEEN THIRTY TO FORTY MINUTES; HOWEVER, THE PATIENT WAS ALREADY HAD GENERAL ANESTHESIA ADMINISTERED, A BOOSTER WAS NOT NECESSARY.

    Patient problems: Foreign Body In Patient

    MAUDE report 21304771 ↗
  • MalfunctionJul 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MPFL SURGERY WITH TWO 4.75 SWIVELOCK, FIBERTAPE AND TIGHTROPE RT THE FIBERTAPE PULLED OUT FROM ONE OF THE SWIVELOCK ANCHORS IN THE PATELLA. THE SURGEON USED LARGER SCREWS IN THE PATELLA, A SCREW IN THE FEMUR AND SYNTHETIC GRAFT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE WITH THE PART NUMBER AR-1555BC. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19740493 ↗
  • MalfunctionAug 8, 2022Break

    IT WAS REPORTED THAT DURING A MEDIAL PATELLOFEMORAL LIGAMENT RENEWAL (MPFL) SURGERY THE ANCHOR BROKE WHEN TIGHTENING THE SUTURES AND TORE OUT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15183116 ↗
  • InjuryAug 24, 2020Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THE PATIENT DEVELOPED AN INFECTION AFTER A MEDIAL PATELLA FEMORAL TENDON WITH ALLOGRAFT PROCEDURE ON (B)(6) 2020. DURING THE (B)(6) 2020 PROCEDURE THE FOLLOWING ARTHREX PRODUCT WAS IMPLANTED: AR-1360C-CP (LOT 10435919). IN ADDITION, A JRF VERSAGRAFT (VRG-OO1) WAS USED. THE PATIENT DEVELOPED AN INFECTION AND ON (B)(6) 2020 THE SURGEON WENT BACK IN AND REMOVED THE IMPLANT AND GRAFT. SURGEON THEN CLEANED AND DRAINED THE WOUND. CULTURES WERE TAKEN AT THE TIME OF SECOND SURGERY AND NO NEW PRODUCTS WERE IMPLANTED. PATIENT WE CONTINUE TO BE OBSERVED DURING THE HEALING PROCESS TO INSURE THE WOUND HEALS AND REMAINS FREE OF INFECTION. THE CULTURE RESULTS WERE POSITIVE FOR (B)(6).

    Patient problems: Bacterial Infection

    MAUDE report 10443132 ↗
  • InjuryMar 19, 2019Break; Detachment of Device or Device Component

    IT WAS REPORTED DURING AN MPFL RECONSTRUCTION PROCEDURE, THE SURGEON WAS USING THE AR-1360C-CP MPFL BIOCOMPOSITE IMPLANT SYSTEM (LOT: 10212210). THE SURGEON IMPLANTED TWO AR-2324BCC SUTURE ANCHOR BIOCOMPOSITE SWIVELOCKS IN THE PATELLA. WHEN THE SURGEON INSERTED THE ANCHOR AND TENODESED THE SEMI-T ALLOGRAFT INTO THE MOST PROXIMAL 4.5MM SOCKET ON THE PATELLA, THE 4.75MM SWIVELOCK INSERTED COMPLETELY, BUT THE MOST PROXIMAL 2-3MM OF THE ANCHOR BROKE AND CAME OFF WHEN THE SWIVELOCK DRIVER WAS REMOVED. THE BROKEN PIECE WAS PRESENT OUTSIDE THE JOINT AND STILL ON THE SWIVELOCK DRIVER, AND WAS EASILY RETRIEVED. THE REP STATED THE REMAINING PORTION OF THE SWIVELOCK ANCHOR WAS SEATED INTO THE SOCKET WITHOUT ISSUE, DESPITE THE PROXIMAL PORTION OF THE ANCHOR BREAKING. THE REST OF THE TECHNIQUE WAS PERFORMED WITHOUT ISSUE. A TOTAL OF THREE IMPLANTS THAT CAME WITHIN THE AR-1360C-CP WERE USED. ONLY THE ONE 4.75MM SWIVELOCK BROKE. THE OTHER SWIVELOCK AND 6X23MM INTERFERENCE SCREW WORKED FINE AND HAD NO ISSUES. NO PUNCH OR TAP WAS USED FOR EITHER SOCKET. THE ONLY DRILL USED FOR EACH SOCKET WAS THE 4.5MM REAMER THAT CAME WITHIN THE KIT. THE QUALITY OF BONE WAS HARD. ADDITIONAL INFORMATION RECEIVED ON

    Patient problems: Failure of Implant

    MAUDE report 8432275 ↗
  • InjuryOct 31, 2017Difficult to Insert; Unstable

    IT WAS REPORTED THAT THE MPFL DRIVER WAS UNAVAILABLE DURING SURGERY BUT THE SURGEON PROCEEDED WITH ANOTHER DRIVER TO INSERT THE SCREW INSTEAD WITH THE SHAFT RETRACTED (AR-1540DB). THE SURGEON STRUGGLED TO INSERT THE SCREW BECAUSE HE ONLY USED THE TIP OF THE DRIVER TO IMPLANT. THE ACUTE ANGLE X-RAY TEMPLATE WAS NOT USED. SCREW WAS INSERTED AND THE PATIENT WAS STABLE IMMEDIATELY POST-OP. A REVISION SURGERY WAS NECESSARY DUE TO INSTABILITY, THE INTERFERENCE SCREW WAS FOUND BROKEN AT PROXIMAL END. THE TIP WILL REMAIN INSIDE THE PATIENT.

    Patient problems: No Code Available

    MAUDE report 6989810 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates