GRYPHON ORTHOCORD 210813Recalls & reported problems

Medos International Sàrl · GRYPHON P BR DS Anchor w/ORTHOCORD TCP/PLGA Absorbable Anchor (1) Violet (1) Blue Strand, Size 2 (5 metric) ORTHOCORD braided composite suture, 36 Inches (91cm) · Class 2 · Implant

All models of GRYPHON ORTHOCORD (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Gryphon OrthoCord suture anchor has 698 adverse event reports, with no deaths and 39 injuries. Many describe the anchor breaking, cracking or pulling out during arthroscopic procedures such as knee and ligament repair, with surgeons removing broken parts and using another device to finish. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
712
Deaths reported
0
Injuries reported
39
Recalls naming it
0

Summary

The FDA's MAUDE database contains 712 adverse event reports that list this exact device (DI 10886705001279), received between Feb 21, 2018 and Sep 30, 2026 — classified as 39 injuries, 673 malfunctions.

The most frequently reported specific device problem is “Break” (509 reports).

7 other recalls cover devices cleared under the same FDA submission (K090124, K150209); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 24 reportsDec 2023: 3 reportsJan 2024: 7 reportsFeb 2024: 7 reportsMar 2024: 8 reportsApr 2024: 8 reportsMay 2024: 3 reportsJun 2024: 17 reportsJul 2024: 4 reportsAug 2024: 9 reportsSep 2024: 7 reportsOct 2024: 7 reportsNov 2024: 11 reportsDec 2024: 10 reportsJan 2025: 13 reportsFeb 2025: 5 reportsMar 2025: 7 reportsApr 2025: 5 reportsMay 2025: 19 reportsJun 2025: 8 reportsJul 2025: 18 reportsAug 2025: 11 reportsSep 2025: 11 reportsOct 2025: 13 reportsNov 2025: 12 reportsDec 2025: 17 reportsJan 2026: 18 reportsFeb 2026: 17 reportsMar 2026: 13 reportsApr 2026: 20 reportsMay 2026: 18 reportsJun 2026: 7 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 25 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break509
  • Migration92
  • Migration or Expulsion of Device25
  • Crack21
  • Detachment of Device or Device Component21
  • Difficult to Advance16
  • Adverse Event Without Identified Device or Use Problem12
  • Device Damaged Prior to Use11

Patient problems

  • No Clinical Signs, Symptoms or Conditions525
  • No Consequences Or Impact To Patient136
  • Injury17
  • Not Applicable13
  • Unspecified Tissue Injury13
  • No Known Impact Or Consequence To Patient8
  • No Code Available7
  • Tissue Damage4

Recent adverse event reports

The 25 most recent of 712 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949210 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948903 ↗
  • MalfunctionSep 29, 2026Migration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933733 ↗
  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923844 ↗
  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923444 ↗
  • MalfunctionSep 28, 2026Break; Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917257 ↗
  • MalfunctionSep 25, 2026Migration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903504 ↗
  • MalfunctionSep 24, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880986 ↗
  • MalfunctionSep 24, 2026Break

    It was reported that during a Rotator Cuff Repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another GRYPHON ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880983 ↗
  • MalfunctionSep 23, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873558 ↗
  • MalfunctionSep 23, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872626 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during an unspecified arthroscopic procedure the GRYPHON P BR DS ANCHOR W/OC anchor device broke on insertion. Another device was used to complete the procedure. There was no delay in the procedure reported. There were no adverse patient consequences reported. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869040 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during a labral repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another two GRYPHON ORTHOCORDs to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 3 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866109 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during a labral repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another two GRYPHON ORTHOCORDs to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866107 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during a labral repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another two GRYPHON ORTHOCORDs to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 1 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866102 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during a Rotator Cuff Repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another GRYPHON ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865714 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that during a Rotator Cuff Repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another GRYPHON ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 1 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865711 ↗
  • MalfunctionSep 22, 2026Break

    It was reported that during a Rotator Cuff Repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another GRYPHON ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857291 ↗
  • MalfunctionSep 22, 2026Break

    It was reported that during a Rotator Cuff Repair surgery, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Changed to another GRYPHON ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 1 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857286 ↗
  • MalfunctionSep 18, 2026Device Contaminated During Manufacture or Shipping

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50821719 ↗
  • MalfunctionSep 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50780671 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that during an unspecified procedure, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769426 ↗
  • MalfunctionSep 11, 2026Break

    It was reported that the event occurred during a knee arthroscopy surgery and it involved two devices. Firstly, when the suture of a RIGIDLOOP device was tightened, the suture broke off. Another device was used to complete the surgery. Second, a GRYPHON ORTHOCORD anchor broke off. All the broken parts were removed. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50729036 ↗
  • MalfunctionSep 9, 2026Break

    It was reported that during a Shoulder arthroscopy procedure, the GRYPHON ORTHOCORD anchor was broken off. All the broken parts were removed. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50688592 ↗
  • MalfunctionSep 1, 2026Migration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50545912 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K090124, K150209. One submission often covers many models, so check the product and lot details in each record.

  • Z-1679-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
  • Z-1678-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
  • Z-1680-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
  • Z-1677-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
  • Z-1681-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
  • Z-2210-2011 · Feb 24, 2011 · Terminated
    DePuy Mitek GRYPHON P BR DS Anchor w/ORTHOCORD Product Code: 210813 Gryphon BR Suture Anchors are intended for: Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Bic
  • Z-2211-2011 · Feb 24, 2011 · Terminated
    DePuy Mitek GRYPHON P BR Anchor w/ORTHOCORD Product Code: 210814 Gryphon BR Suture Anchors are intended for: Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Bicep

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →