Arthrex® AR-2324BCCRecalls & reported problems
Arthrex, Inc. · BIO-COMP SWVLK C, CLD 4.75X19.1MM · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex BioComposite SwiveLock anchor has 248 reports on file, with no deaths and 36 injuries. Many describe the anchor or its screw breaking during insertion or fixation, some involving broken fragments, and others describe anchors that could not be fixed in bone. No recalls name this device, and reports fell to 35 in the latest year from 71 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 247
- Deaths reported
- 0
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 247 adverse event reports that list this exact device (DI 00888867026827), received between Aug 23, 2016 and Sep 17, 2026 — classified as 36 injuries, 211 malfunctions.
The most frequently reported specific device problem is “Break” (157 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break157
- Device Dislodged or Dislocated21
- Failure to Advance11
- Detachment of Device or Device Component8
- Patient-Device Incompatibility8
- Material Deformation7
- Physical Resistance/Sticking7
- Defective Device5
Patient problems
- No Clinical Signs, Symptoms or Conditions159
- Insufficient Information54
- Foreign Body In Patient11
- Failure of Implant10
- Pain9
- Swelling5
- Hypersensitivity/Allergic reaction3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 25 most recent of 247 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Defective Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50804960 ↗MalfunctionAug 4, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50132855 ↗MalfunctionJul 29, 2026Break
It was reported that during an posterior cruciate ligament reconstruction with internal brace the screw broke in half when the surgeon removed the swivelock during fixation of the internal brace in the tibia. All fragments were retrieved. Per complaint reporter there was no harm for patient, operator or third party. The surgery was finished successfully with a new device with the same part number. It was not necessary to switch the surgical technique or do a second surgery.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50073090 ↗MalfunctionJun 5, 2026Positioning Failure
IT WAS REPORTED THAT DURING A QUADRICEPS TENDON REFIXATION SURGERY 4 DEVICES COULD NOT BE FIXATED IN THE BONE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25406815 ↗MalfunctionApr 1, 2026Break
ON 12TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24752899 ↗MalfunctionMar 30, 2026Break
ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24715272 ↗MalfunctionMar 30, 2026Break
ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24714443 ↗MalfunctionMar 26, 2026Break
ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING AN ANKLE JOINT INSTABILITY SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24698348 ↗MalfunctionMar 25, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A SURGERY THE ANCHOR WAS IMPOSSIBLE TO SEPARATE FROM THE DRIVER AFTER INSERTION. THE SURGEONS TRIED TWO MORE TIMES WITHOUT SUCCESS AND THEN OPENED A NEW DEVICE WHICH WORKED FINE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24688774 ↗MalfunctionMar 25, 2026Positioning Failure
ON 5TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, THE ANCHOR COULD NOT BE SET SUCCESSFULLY. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS ANOTHER NEW AR-2324BCC WAS USED. AR-1927PB AND AR-2324PTB WERE USED TO PREPARE THE BONE SOCKET. THE BONE QUALITY OF PATIENT WAS SOFT AND NO FRAGMENTS WERE FOUND. 13TH MARCH 2026 - THE COMPLAINT INITIATOR REPLIED THAT THE ANCHOR COULD NOT BE FULLY IMPLANTED, SO THE SURGEON NEEDED TO REPLACE A NEW ANCHOR TO COMPLETE THE PROCEDURE. HE CONFIRMED THAT THE ANCHOR WAS DEPLOYED SUCCESSFULLY, BUT IMPLANT WOULD NOT SEAT.
Patient problems: Insufficient Information
MAUDE report 24688393 ↗MalfunctionMar 25, 2026Material Deformation
ON 5TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET; THE ANCHOR WAS DEFORMED DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS ANOTHER NEW AR-2324BCC WAS USED. AR-1927PB AND AR-2324PTB WERE USED TO PREPARE BONE SOCKET. THE BONE QUALITY OF PATIENT WAS HARD, AND NO FRAGMENTS WERE LEFT IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 24687915 ↗MalfunctionMar 25, 2026Break
ON 5TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS ANOTHER NEW AR-2324BCC WAS USED. AR-1927PB AND AR-2324PTB WERE USED TO PREPARE BONE SOCKET AND THE BONE QUALITY OF PATIENT WAS HARD. THERE WERE NO FRAGMENTS LEFT IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 24687516 ↗MalfunctionMar 23, 2026Device Dislodged or Dislocated
ON 4TH MAR 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-13995N CORPION-MULTIFIRE NEEDLE BROKE AFTER THE SUTURE WAS PASSED THROUGH. IN ADDITION, THE BIO-COMP SWVLK C, CLD 4.75 X 19.1 MM, BATCH 15412798, WAS OBSERVED TO HAVE SWIVELOCK ANCHOR PULLED OUT. THIS WAS DETECTED DURING THE PROCEDURE ON (B)(6) 2026 AND WAS POSSIBLY DUE TO THE PATIENT HAVING SOFT BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24661375 ↗MalfunctionMar 23, 2026Break
ON 3RD MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660999 ↗MalfunctionMar 23, 2026Break
ON 3RD MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, BOTH ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660845 ↗MalfunctionMar 23, 2026Break
ON 3RD MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, BOTH ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660774 ↗MalfunctionMar 18, 2026Device Dislodged or Dislocated
ON 27TH FEB 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR PULLED OUT DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24628464 ↗MalfunctionMar 3, 2026Break
ON 11-FEB-2026, A DISTRIBUTOR REPORTED VIA EMAIL THAT AN AR-2324BCC BIOCOMPOSITE SWIVELOCK C ANCHOR BROKE DURING INSERTION OF THE SWIVELOCK. THE ISSUE WAS IDENTIFIED INTRAOPERATIVELY ON (B)(6) 2026, AND NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 19-FEB-2026: DURING A SUBSCAPULARIS REPAIR PROCEDURE, THE THREADS OF AN AR-2324BCC BIOCOMPOSITE SWIVELOCK C ANCHOR STRIPPED UPON INSERTION. ALL ASSOCIATED FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THE SURGEON REPLACED THE DEVICE WITH AN AR-2324BCC IMPLANT TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED DELAY IN THE CASE OR ADDITIONAL ANESTHESIA TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24499655 ↗MalfunctionFeb 25, 2026Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY THE EYELET OF THE ANCHOR BROKE IN HALF WHEN THE ANCHOR WAS PLACED IN THE HOLE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-2323PSLM). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442460 ↗MalfunctionFeb 19, 2026Break
ON 03-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT TWO AR-2324BCC BIOCOMPOSITE SWIVELOCK C ANCHORS HAD THE TIPS BREAK DURING INSERTION. THE PROCEDURE WAS COMPLETED USING A REPLACEMENT AR-2324BCC BIOCOMPOSITE SWIVELOCK C. NO FRAGMENTS WERE RETAINED IN THE PATIENT. THERE WAS NO SIGNIFICANT DELAY, NO ADDITIONAL ANESTHESIA WAS ADMINISTERED, AND NO ADVERSE EFFECTS WERE REPORTED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2026 WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24392221 ↗MalfunctionFeb 19, 2026Break
ON 03-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT TWO AR-2324BCC BIOCOMPOSITE SWIVELOCK C ANCHORS HAD THE TIPS BREAK DURING INSERTION. THE PROCEDURE WAS COMPLETED USING A REPLACEMENT AR-2324BCC BIOCOMPOSITE SWIVELOCK C. NO FRAGMENTS WERE RETAINED IN THE PATIENT. THERE WAS NO SIGNIFICANT DELAY, NO ADDITIONAL ANESTHESIA WAS ADMINISTERED, AND NO ADVERSE EFFECTS WERE REPORTED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2026, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24391193 ↗MalfunctionFeb 18, 2026Break
ON 29-JAN-2026, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT THE TIP OF AN AR-2324BCC BIOCOMPOSITE SWIVELOCK INSERTER BROKE WHILE INSERTING THE DEVICE INTO THE BONE AFTER A PILOT HOLE WAS CREATED. THE SURGEON WAS UNABLE TO REMOVE SEVERAL OF THE BROKEN FRAGMENTS, WHICH REMAIN IN THE PATIENT. THERE ARE NO CURRENT PLANS FOR A SECOND PROCEDURE TO REMOVE THE FRAGMENTS. NO SIGNIFICANT SURGICAL DELAY WAS NOTED, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THIS ISSUE WAS DISCOVERED DURING AN ANTERIOR CAPSULAR LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2026. THERE WAS NO PROBLEM WITH THE STRENGTH OF THE GRAFT FIXATION PROVIDED BY THE PRODUCT, SO THE SURGERY WAS COMPLETED WITHOUT FURTHER ISSUES.
Patient problems: Foreign Body In Patient
MAUDE report 24376190 ↗MalfunctionFeb 13, 2026Break
ON 27TH JAN 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, IT ANCHORS BROKE DURING INSERTION. THIS WAS DETECTED DURING ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24331763 ↗MalfunctionFeb 4, 2026Break
ON 15TH JAN 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-2324BCC, SUTURE ANCHOR, BIOCOMPOSITE SWIVELOCK® C, 4.75 MM X 19.1 MM, CLOSED EYELET, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-2324BCC. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259930 ↗MalfunctionJan 22, 2026Break
ON 7TH JANUARY 2026, A DISTRIBUTOR REPORTED VIA EMAIL THAT THE TIP OF THE SCREW FROM AR-2324BCC, LOT 15491447 TIP OF THE SCREW FRACTURED DURING INSERTION IN A SHOULDER PROCEDURE PERFORMED ON (B)(6) 2025. THE SURGERY WAS COMPLETED WITHOUT COMPLICATIONS USING A NEW AR-2324BCC, WITH NO FRAGMENTS RETAINED IN THE PATIENT¿S BODY AND NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24143843 ↗
FDA clearance and classification
The FDA regulates product code MAI (fastener, fixation, biodegradable, soft tissue) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Arthrex® AR-2324BCC
Has the Arthrex® AR-2324BCC been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888867026827), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Arthrex® AR-2324BCC?
The FDA's MAUDE database lists 247 adverse event reports for this device, received between Aug 23, 2016 and Sep 17, 2026. In those reports, the most common specific device problems are “Break” (157), “Device Dislodged or Dislocated” (21) and “Failure to Advance” (11); the most common patient problems are “Foreign Body In Patient” (11), “Failure of Implant” (10) and “Pain” (9). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Arthrex® AR-2324BCC?
None of the reports are classified as deaths, 36 as injuries and 211 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Arthrex® AR-2324BCC MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the Arthrex® AR-2324BCC, and is it still sold?
The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-2324BCC). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE688 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall586 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE585 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue419 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE319 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE284 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall191 reports
- Arthrex® AR-1927BCF-45Arthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE155 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.