Skip to content

Arthrex® AR-1380BRecalls & reported problems

Arthrex, Inc. · BIO-INTERFERENCE SCREW · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867015265), received between Jun 11, 2024 and Aug 28, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Damaged Thread” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Damaged Thread1
  • Material Deformation1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2024Material Deformation; Material Deformation

    ON 08/07/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-1380B BIO-INTERFERENCE SCREW WITH DISPOSABLE SHEATH DOES NOT SCREW IN THE PATIENT'S TENDON. THE SURGEON APPLIED FORCE TO INSERT THE SCREW, BUT IT DID NOT SCREW IN. WHEN THE SCREW WAS REMOVED, IT APPEARED DEFORMED. THIS WAS DISCOVERED DURING A KNEE ARTHROSCOPY AND ACL PROCEDURE ON (B)(6) 2024. ADDITIONAL INFORMATION RECEIVED ON 8/26/2024: UPON REMOVAL OF THE SCREW, IT WAS DEFORMED; HOWEVER, IT DID NOT BREAK IN THE PATIENT. THE ENTIRE SCREW WAS SUCCESSFULLY REMOVED. THE CASE WAS COMPLETED USING AN AR-4030C-09 BIOCOMPOSITE FASTTHREAD INTERFERENCE SCREW. THE CASE WAS DELAYED TEN MINUTES WITHOUT ADDITIONAL ANESTHESIA.

    Patient problems: Insufficient Information

    MAUDE report 20087194 ↗
  • MalfunctionJun 11, 2024Damaged Thread; Damaged Thread

    ON 4/4/2023 IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-1380B BIO-INTERFERENCE SCREW WITH DISPOSABLE SHEATH HAD AN ISSUE. AR-4019D-1 FASTTHREAD SCREW DRIVER FOR 6 MM AND AR-1996CD-1 QUICK CONNECT DRIVER WERE USED TO ANCHOR THE SCREW FOR INSERTION. HOWEVER, WHEN THE SURGEON WAS SCREWING IT IN, THE SCREW DID NOT THREAD ANYMORE. THIS WAS DISCOVERED DURING AN LCA PROCEDURE WITH NO PATIENT HARM. ADDITIONAL INFORMATION IS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19510994 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates