Arthrex® AR-1380TRecalls & reported problems
Arthrex, Inc. · SCREW, CANN. INT., FULL THREAD · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867015333), received between May 17, 2018 and May 17, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 17, 2018Break
IT WAS REPORTED THAT DURING AN ACL REVISION (WHICH INVOLVED REMOVAL OF TITANIUM INTERFERENCE SCREWS FROM A PREVIOUS SURGERY) AND A MENISCAL ROOT REPAIR PROCEDURE, THE SURGEON PREPARED THE BTB ALLOGRAFT WITH #5 FIBERWIRE AND #2 FIBERWIRE. THE FEMORAL SOCKET WAS PREPARED THROUGH THE MEDIAL PORTAL WITH THE OVER-THE-TOP GUIDE AND 2.4 MM BEATH PIN, FOLLOWED BY A 10 MM LOW PROFILE REAMER TO A DEPTH OF 30 MM. THE TIBIAL TUNNEL WAS PREPARED WITH AN ACUFEX GUIDE, 2.4 MM DRILL TIP PIN, AND 10 MM CANNULATED DRILL. THE GRAFT WAS PASSED WITH A #2 FIBERWIRE THROUGH THE TIBIAL TUNNEL AND INTO THE FEMORAL SOCKET. AN 8 MM TAP WITH DISPOSABLE SHEATH AND TUNNEL NOTCHER WAS USED ON THE FEMORAL SOCKET. AN 8 X 23 MM BIOCOMPOSITE INTERFERENCE SCREW W/ DISPOSABLE SHEATH (AR-1380BC, LOT: 10051207, (B)(4)) WAS INSERTED AND BROKE WHILE THE SURGEON WAS INSERTING IT. ALL FRAGMENTS OF THIS SCREW WERE REMOVED AND A 7 MM TAP WAS USED TO RE-TAP THE FEMORAL SOCKET. A SECOND SCREW, A 7 X 23 MM BIOCOMPOSITE INTERFERENCE SCREW W/ DISPOSABLE SHEATH (AR-1370C, LOT: 10192231, (B)(4)) WAS INSERTED AND THIS SCREW BROKE AS WELL AS THE SURGEON WAS INSERTING IT. ALL FRAGMENTS OF THIS SCREW WERE REMOVED. AT THIS POINT A DIFFER
Patient problems: No Code Available
MAUDE report 7521508 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page