Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867015340), received between Oct 10, 2017 and Oct 18, 2017 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Difficult to Insert2
- Difficult or Delayed Positioning1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
Patient problems
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 18, 2017Break; Difficult to Insert; Difficult to Remove; Improper or Incorrect Procedure or Method
IT WAS ORIGINALLY REPORTED THAT DURING THE SURGERY THE SCREW BROKE. INVESTIGATION UPDATE: ONE HALF REMAINED INSIDE THE PATIENT THE OTHER HALF WAS RETRIEVED. THE SCREW DOES NOT ENTER EASILY INTO THE BONE, BUT THE SURGEON TRIED ANYWAY (WITH FORCE) AND SO THE SCREW BROKE ON INSERTION.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6958023 ↗InjuryOct 18, 2017Break; Difficult or Delayed Positioning
IT WAS ORIGINALLY REPORTED THAT DURING THE SURGERY THE SCREW BROKE. INVESTIGATION UPDATE: ONE HALF REMAINED INSIDE THE PATIENT THE OTHER HALF WAS RETRIEVED. THE SCREW DOES NOT ENTER EASILY INTO THE BONE, BUT THE SURGEON TRIED ANYWAY (WITH FORCE) AND SO THE SCREW BROKE ON INSERTION.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6958012 ↗InjuryOct 10, 2017Break; Difficult to Insert
IT WAS ORIGINALLY REPORTED THAT THERE WAS A THREAD DEFECT AND THE SCREW BROKE. THE SURGEON WAS ABLE TO REMOVE THE BROKEN FRAGMENT. THE OTHER HALF REMAINED INSIDE THE PATIENT. THE SURGEON CONFIRMED THAT THE TENDON IS SECURELY FIXED IN THE TUNNEL. FOLLOW UP INVESTIGATION: THE SCREW DOES NOT ENTER EASILY INTO THE BONE, BUT THE SURGEON TRIED ANYWAY (WITH FORCE) AND SO THE SCREW BROKE ON INSERTION.
Patient problems: Foreign Body In Patient
MAUDE report 6934416 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page