Arthrex® AR-1386Recalls & reported problems
Arthrex, Inc. · SCRWDRV,CANN FOR BIO INT. SCRW · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867015425), received between Feb 21, 2018 and Dec 26, 2019 — classified as 5 injuries.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break5
- Difficult to Remove2
Patient problems
- Device Embedded In Tissue or Plaque3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 26, 2019Break; Difficult to Remove
IT WAS REPORTED THAT THE ARTHREX AR-1386 SCREWDRIVER WAS USED DURING ANTERIOR CRUCIATE LIGAMENT SURGERY. DURING INSERTION OF THE SCREW, THE TIP OF THE SCREWDRIVER HAS BROKEN OFF INSIDE THE SCREW. IT WAS NOT POSSIBLE FOR THE SURGEON TO REMOVE THE TIP. THE PATIENT IS VERY YOUNG. THE PARENTS AND THE PATIENT WERE INFORMED ABOUT THE INCIDENT. THE SURGERY WAS STILL COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION RECEIVED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9519620 ↗InjuryOct 1, 2019Break
IT WAS REPORTED THAT THE TIP OF SCREWDRIVER BROKE INSIDE THE SCREW DURING IMPLANTATION. UPDATE 10-SEPT-2019: AN ANTERIOR CRUCIATE LIGAMENT (ACL) PROCEDURE WAS PERFORMED. THE BROKEN PIECE REMAINED IN THE SCREW INSIDE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9141367 ↗InjuryJan 30, 2019Break
IT WAS REPORTED THAT DURING A ANTERIOR CRUCIATE LIGAMENT SURGERY, WHEN SCREWING THE FEMORAL SCREW, THE SCREWDRIVER BROKE. 1 FRAGMENT REMAINED IN THE FEMORAL SCREW AND IT IS NOT POSSIBLE TO REMOVE IT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WHICH WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 10TH. JAN. 2019: THE BROKEN PIECE IS SET UP INSIDE THE SCREW, WELL IMPLANTED INTO THE BONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8288790 ↗InjuryJun 7, 2018Break
IT WAS REPORTED THAT THE TIP OF THE SCREWDRIVER BROKE INSIDE THE BIOABSORBABLE SCREW. THE BROKEN FRAGMENT WAS NOT REMOVED BUT REMAINED INSIDE THE SCREW AND WAS FIXED THERE. THE SURGERY WAS COMPLETED SUCCESSFULLY, NO NEGATIVE CONSEQUENCES FOR THE PATIENT HAVE BEEN REPORTED SO FAR. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. FOLLOW-UP INVESTIGATION: AN X-RAY WAS DONE, BUT NO COMPUTERIZED PICTURE IS AVAILABLE. THE TRANSPLANT IS BLOCKED IN THE FEMUR TUNNEL WITH THE SCREW. THE SURGEON WAS NOT ABLE TO GIVE MORE INFORMATION. THE PART NUMBER OF THE SCREW IS AR-1370B WITH BATCH NO. 10169107.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7577715 ↗InjuryFeb 21, 2018Break; Difficult to Remove
IT WAS REPORTED THAT THE AR-1380B BROKE, THEN THE AR-1390B, THEN THE SCREWDRIVER BROKE. THE TIP OF THE SCREW DRIVER REMAINED IN THE PATIENT. THE FRAGMENTS OF THE SCREWS WERE REMOVED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7285128 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page