Skip to content

Arthrex® AR-1400TCRecalls & reported problems

Arthrex, Inc. · BIO-COMPR INTERF SCRW,FULL THD · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867016231), received between Jan 31, 2022 and Oct 12, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 12, 2023Break; Break

    FURTHER INFORMATION WERE PROVIDED THAT THE SURGEON TRIED TO IMPLANT IN THE SCREW INTO THE BONE BUT THE SCREW DIDN¿T ADVANCE AND THE THREAD DIDN¿T GRIP, WHILE HE FORCED THE IMPLANT THE TIP OF THE SCREW BROKE OFF BUT IT WAS AT THE BEGINNING OF THE TIBIAL TUNNEL SO THE SURGEON WAS ABLE TO REMOVE THE FRAGMENT WITH A CLAMP.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17919824 ↗
  • MalfunctionApr 26, 2023Break; Break

    ON 03/28/2023, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA SEMS THAT AN AR-1400TC BIOCOMPOSITE INTERFERENCE SCREW BROKE. THIS OCCURRED DURING A CASE, DURING INSERTION THE SCREW WAS NOT PROGRESSING THE SURGEON INSISTED UNTIL THE SCREW BROKE AND THE KEY BENT. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16815007 ↗
  • MalfunctionJan 31, 2022Break; Output Problem

    ON (B)(6) 2022, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-1400TC FULL THREAD CANNULATED BIOCOMPOSITE INTERFERENCE SCREW GOES IN WELL INTO THE TUNNEL; HOWEVER, WHEN REMOVING THE SCREWDRIVER, THE SCREW CAME OUT WITH IT, DAMAGED, AND WAS NOT ABLE TO USE AGAIN. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 13393090 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page