Arthrex® AR-1407Recalls & reported problems
Arthrex, Inc. · HEADED REAMER 7MM, CANNULATED · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867016668), received between May 8, 2019 and Jun 18, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Dull, Blunt1
- Entrapment of Device1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2024Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT THE DEVICE IS DULL. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19563895 ↗InjuryMay 8, 2019Break; Entrapment of Device
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) PROCEDURE THAT AN AR-1407 (7MM CANNULATED HEADED REAMER) WAS USED. IT WAS INDICATED THAT MATERIAL BROKE OFF FROM THE INSTRUMENT AND REMAINS IN THE PATIENT. THE SURGERY WAS FINISHED WITH A NEW DEVICE WITH A DIFFERENT PART NUMBER, AR-1408. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 18-APR-2019: ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT PART AR-1887-08P (TRANSFIX TUNNEL HOOK 7) WAS ALSO INVOLVED IN THIS INCIDENT. THIS DEVICE IS NOT APPROVED FOR SALE IN THE USA. UPDATE 29-APR-2019: THE HEAD OF THE DEVICE AR-1887-08P, WHICH ACTUALLY SHOULD PULL THE WIRE OUT, DID NOT COME OUT, HENCE IT BROKE OFF. IT WAS A DEVICE WHICH HAS BEEN USED VERY RARE AND WHICH HAS BEEN CONTROLLED ROUTINELY. SINCE THIS INCIDENT SENSITIVE CONTROLS HAVE BEEN INITIATED. A DROPPING OR ANY OTHER MISUSE OF THE AFFECTED DEVICE CANNOT BE DETECTED. THE LOOP OF THIS DEVICE REMAINED INSIDE THE PATIENT. IT WAS NOTED THAT IT CAN ONLY BEEN REMOVED WITH HIGH TECHNICAL EFFORT AND THAT LEAVING IT REMAINING INSIDE THE PATIENT CAN BE EXPECTED TO HAVE NO FURTHER CONSEQUENCES. HENCE THE PART HAS REMAINED INSIDE THE PATIENT. AF
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8591600 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page